se ha leído el artículo
array:24 [ "pii" => "S0014256524000638" "issn" => "00142565" "doi" => "10.1016/j.rce.2024.03.001" "estado" => "S300" "fechaPublicacion" => "2024-05-01" "aid" => "2185" "copyright" => "Elsevier España, S.L.U. and Sociedad Española de Medicina Interna (SEMI)" "copyrightAnyo" => "2024" "documento" => "article" "crossmark" => 1 "subdocumento" => "rev" "cita" => "Rev Clin Esp. 2024;224:314-30" "abierto" => array:3 [ "ES" => false "ES2" => false "LATM" => false ] "gratuito" => false "lecturas" => array:1 [ "total" => 0 ] "Traduccion" => array:1 [ "en" => array:19 [ "pii" => "S2254887424000523" "issn" => "22548874" "doi" => "10.1016/j.rceng.2024.04.005" "estado" => "S300" "fechaPublicacion" => "2024-05-01" "aid" => "2185" "copyright" => "Elsevier España, S.L.U. and Sociedad Española de Medicina Interna (SEMI)" "documento" => "article" "crossmark" => 1 "subdocumento" => "rev" "cita" => "Rev Clin Esp. 2024;224:314-30" "abierto" => array:3 [ "ES" => false "ES2" => false "LATM" => false ] "gratuito" => false "lecturas" => array:1 [ "total" => 0 ] "en" => array:13 [ "idiomaDefecto" => true "cabecera" => "<span class="elsevierStyleTextfn">Review</span>" "titulo" => "Determining factors of pulse oximetry accuracy: a literature review" "tienePdf" => "en" "tieneTextoCompleto" => "en" "tieneResumen" => array:2 [ 0 => "en" 1 => "es" ] "paginas" => array:1 [ 0 => array:2 [ "paginaInicial" => "314" "paginaFinal" => "330" ] ] "titulosAlternativos" => array:1 [ "es" => array:1 [ "titulo" => "Factores determinantes de la precisión de la oximetría de pulso: revisión bibliográfica" ] ] "contieneResumen" => array:2 [ "en" => true "es" => true ] "contieneTextoCompleto" => array:1 [ "en" => true ] "contienePdf" => array:1 [ "en" => true ] "resumenGrafico" => array:2 [ "original" => 0 "multimedia" => array:8 [ "identificador" => "fig0005" "etiqueta" => "Figure 1" "tipo" => "MULTIMEDIAFIGURA" "mostrarFloat" => true "mostrarDisplay" => false "figura" => array:1 [ 0 => array:4 [ "imagen" => "gr1.jpeg" "Alto" => 1343 "Ancho" => 1667 "Tamanyo" => 152618 ] ] "detalles" => array:1 [ 0 => array:3 [ "identificador" => "at0005" "detalle" => "Figure " "rol" => "short" ] ] "descripcion" => array:1 [ "en" => "<p id="spar0005" class="elsevierStyleSimplePara elsevierViewall">Flow chart consisting of the statement of preferred reporting items for systematic reviews (PRISMA).</p>" ] ] ] "autores" => array:1 [ 0 => array:2 [ "autoresLista" => "D. León-Valladares, L.A. Barrio-Mateu, N. Cortés-Carmona, G. Fuentes-Lizana, A.M. Cabanas, K. Latorre-Progulakis, M. Fuentes-Guajardo, P. Martín-Escudero" "autores" => array:8 [ 0 => array:2 [ "nombre" => "D." "apellidos" => "León-Valladares" ] 1 => array:2 [ "nombre" => "L.A." "apellidos" => "Barrio-Mateu" ] 2 => array:2 [ "nombre" => "N." "apellidos" => "Cortés-Carmona" ] 3 => array:2 [ "nombre" => "G." "apellidos" => "Fuentes-Lizana" ] 4 => array:2 [ "nombre" => "A.M." "apellidos" => "Cabanas" ] 5 => array:2 [ "nombre" => "K." "apellidos" => "Latorre-Progulakis" ] 6 => array:2 [ "nombre" => "M." "apellidos" => "Fuentes-Guajardo" ] 7 => array:2 [ "nombre" => "P." "apellidos" => "Martín-Escudero" ] ] ] ] ] "idiomaDefecto" => "en" "Traduccion" => array:1 [ "es" => array:9 [ "pii" => "S0014256524000638" "doi" => "10.1016/j.rce.2024.03.001" "estado" => "S300" "subdocumento" => "" "abierto" => array:3 [ "ES" => false "ES2" => false "LATM" => false ] "gratuito" => false "lecturas" => array:1 [ "total" => 0 ] "idiomaDefecto" => "es" "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S0014256524000638?idApp=WRCEE" ] ] "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S2254887424000523?idApp=WRCEE" "url" => "/22548874/0000022400000005/v3_202405260719/S2254887424000523/v3_202405260719/en/main.assets" ] ] "itemSiguiente" => array:19 [ "pii" => "S001425652400064X" "issn" => "00142565" "doi" => "10.1016/j.rce.2024.02.007" "estado" => "S300" "fechaPublicacion" => "2024-05-01" "aid" => "2186" "copyright" => "Elsevier España, S.L.U. and Sociedad Española de Medicina Interna (SEMI)" "documento" => "simple-article" "crossmark" => 1 "subdocumento" => "cor" "cita" => "Rev Clin Esp. 2024;224:331-2" "abierto" => array:3 [ "ES" => false "ES2" => false "LATM" => false ] "gratuito" => false "lecturas" => array:1 [ "total" => 0 ] "es" => array:11 [ "idiomaDefecto" => true "cabecera" => "<span class="elsevierStyleTextfn">CORRESPONDENCIA</span>" "titulo" => "Carta al editor sobre «Los retos de la bioética en la práctica clínica del siglo XXI»" "tienePdf" => "es" "tieneTextoCompleto" => "es" "paginas" => array:1 [ 0 => array:2 [ "paginaInicial" => "331" "paginaFinal" => "332" ] ] "titulosAlternativos" => array:1 [ "en" => array:1 [ "titulo" => "Letter to the editor on «Bioethical challenges in clinical practice of the 20th century»" ] ] "contieneTextoCompleto" => array:1 [ "es" => true ] "contienePdf" => array:1 [ "es" => true ] "resumenGrafico" => array:2 [ "original" => 0 "multimedia" => array:7 [ "identificador" => "fig0005" "etiqueta" => "Figura 1" "tipo" => "MULTIMEDIAFIGURA" "mostrarFloat" => true "mostrarDisplay" => false "figura" => array:1 [ 0 => array:4 [ "imagen" => "gr1.jpeg" "Alto" => 3482 "Ancho" => 1675 "Tamanyo" => 282442 ] ] "descripcion" => array:1 [ "es" => "<p id="spar0005" class="elsevierStyleSimplePara elsevierViewall">Normas éticas internacionales. ONU: Organización de las Naciones Unidas.</p>" ] ] ] "autores" => array:1 [ 0 => array:2 [ "autoresLista" => "M.F. García Aguilera, N. García Méndez, E. Álvarez Muñoz, J. Lagarda Cuevas" "autores" => array:4 [ 0 => array:2 [ "nombre" => "M.F." "apellidos" => "García Aguilera" ] 1 => array:2 [ "nombre" => "N." "apellidos" => "García Méndez" ] 2 => array:2 [ "nombre" => "E." "apellidos" => "Álvarez Muñoz" ] 3 => array:2 [ "nombre" => "J." "apellidos" => "Lagarda Cuevas" ] ] ] ] ] "idiomaDefecto" => "es" "Traduccion" => array:1 [ "en" => array:9 [ "pii" => "S225488742400047X" "doi" => "10.1016/j.rceng.2024.03.004" "estado" => "S300" "subdocumento" => "" "abierto" => array:3 [ "ES" => false "ES2" => false "LATM" => false ] "gratuito" => false "lecturas" => array:1 [ "total" => 0 ] "idiomaDefecto" => "en" "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S225488742400047X?idApp=WRCEE" ] ] "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S001425652400064X?idApp=WRCEE" "url" => "/00142565/0000022400000005/v2_202405080641/S001425652400064X/v2_202405080641/es/main.assets" ] "itemAnterior" => array:19 [ "pii" => "S0014256524000663" "issn" => "00142565" "doi" => "10.1016/j.rce.2024.03.003" "estado" => "S300" "fechaPublicacion" => "2024-05-01" "aid" => "2188" "copyright" => "Elsevier España, S.L.U. and Sociedad Española de Medicina Interna (SEMI)" "documento" => "article" "crossmark" => 1 "subdocumento" => "fla" "cita" => "Rev Clin Esp. 2024;224:300-13" "abierto" => array:3 [ "ES" => false "ES2" => false "LATM" => false ] "gratuito" => false "lecturas" => array:1 [ "total" => 0 ] "es" => array:13 [ "idiomaDefecto" => true "cabecera" => "<span class="elsevierStyleTextfn">Artículo especial</span>" "titulo" => "Diagnóstico y tratamiento de la trombosis venosa profunda de las extremidades inferiores y superiores. Recomendaciones del grupo de enfermedad tromboembólica de la Sociedad Española de Medicina Interna 2024" "tienePdf" => "es" "tieneTextoCompleto" => "es" "tieneResumen" => array:2 [ 0 => "es" 1 => "en" ] "paginas" => array:1 [ 0 => array:2 [ "paginaInicial" => "300" "paginaFinal" => "313" ] ] "titulosAlternativos" => array:1 [ "en" => array:1 [ "titulo" => "Diagnosis and treatment of deep vein thrombosis of the lower and upper limbs. 2024 recommendations of the thromboembolic disease group of the Spanish Society of Internal Medicine" ] ] "contieneResumen" => array:2 [ "es" => true "en" => true ] "contieneTextoCompleto" => array:1 [ "es" => true ] "contienePdf" => array:1 [ "es" => true ] "resumenGrafico" => array:2 [ "original" => 0 "multimedia" => array:7 [ "identificador" => "fig0005" "etiqueta" => "Figura 1" "tipo" => "MULTIMEDIAFIGURA" "mostrarFloat" => true "mostrarDisplay" => false "figura" => array:1 [ 0 => array:4 [ "imagen" => "gr1.jpeg" "Alto" => 1796 "Ancho" => 2091 "Tamanyo" => 242325 ] ] "descripcion" => array:1 [ "es" => "<p id="spar0020" class="elsevierStyleSimplePara elsevierViewall">Sistema venoso superficial y profundo de los MMII.</p>" ] ] ] "autores" => array:1 [ 0 => array:2 [ "autoresLista" => "A. Gil Díaz, J. Martín Guerra, P. Parra Caballero, G. Puche Palao, N. Muñoz Rivas, N. Ruiz-Giménez Arrieta" "autores" => array:6 [ 0 => array:2 [ "nombre" => "A." "apellidos" => "Gil Díaz" ] 1 => array:2 [ "nombre" => "J." "apellidos" => "Martín Guerra" ] 2 => array:2 [ "nombre" => "P." "apellidos" => "Parra Caballero" ] 3 => array:2 [ "nombre" => "G." "apellidos" => "Puche Palao" ] 4 => array:2 [ "nombre" => "N." "apellidos" => "Muñoz Rivas" ] 5 => array:2 [ "nombre" => "N." "apellidos" => "Ruiz-Giménez Arrieta" ] ] ] ] ] "idiomaDefecto" => "es" "Traduccion" => array:1 [ "en" => array:9 [ "pii" => "S2254887424000511" "doi" => "10.1016/j.rceng.2024.04.004" "estado" => "S300" "subdocumento" => "" "abierto" => array:3 [ "ES" => false "ES2" => false "LATM" => false ] "gratuito" => false "lecturas" => array:1 [ "total" => 0 ] "idiomaDefecto" => "en" "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S2254887424000511?idApp=WRCEE" ] ] "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S0014256524000663?idApp=WRCEE" "url" => "/00142565/0000022400000005/v2_202405080641/S0014256524000663/v2_202405080641/es/main.assets" ] "es" => array:20 [ "idiomaDefecto" => true "cabecera" => "<span class="elsevierStyleTextfn">REVISIÓN</span>" "titulo" => "Factores determinantes de la precisión de la oximetría de pulso: revisión bibliográfica" "tieneTextoCompleto" => true "paginas" => array:1 [ 0 => array:2 [ "paginaInicial" => "314" "paginaFinal" => "330" ] ] "autores" => array:1 [ 0 => array:4 [ "autoresLista" => "D. León-Valladares, L.A. Barrio-Mateu, N. Cortés-Carmona, G. Fuentes-Lizana, A.M. Cabanas, K. Latorre-Progulakis, M. Fuentes Guajardo, P. Martín-Escudero" "autores" => array:8 [ 0 => array:4 [ "nombre" => "D." "apellidos" => "León-Valladares" "email" => array:1 [ 0 => "daynerileon1@gmail.com" ] "referencia" => array:2 [ 0 => array:2 [ "etiqueta" => "<span class="elsevierStyleSup">a</span>" "identificador" => "aff0005" ] 1 => array:2 [ "etiqueta" => "<span class="elsevierStyleSup">*</span>" "identificador" => "cor0005" ] ] ] 1 => array:3 [ "nombre" => "L.A." "apellidos" => "Barrio-Mateu" "referencia" => array:1 [ 0 => array:2 [ "etiqueta" => "<span class="elsevierStyleSup">a</span>" "identificador" => "aff0005" ] ] ] 2 => array:3 [ "nombre" => "N." "apellidos" => "Cortés-Carmona" "referencia" => array:1 [ 0 => array:2 [ "etiqueta" => "<span class="elsevierStyleSup">b</span>" "identificador" => "aff0010" ] ] ] 3 => array:3 [ "nombre" => "G." "apellidos" => "Fuentes-Lizana" "referencia" => array:1 [ 0 => array:2 [ "etiqueta" => "<span class="elsevierStyleSup">b</span>" "identificador" => "aff0010" ] ] ] 4 => array:3 [ "nombre" => "A.M." "apellidos" => "Cabanas" "referencia" => array:1 [ 0 => array:2 [ "etiqueta" => "<span class="elsevierStyleSup">c</span>" "identificador" => "aff0015" ] ] ] 5 => array:3 [ "nombre" => "K." "apellidos" => "Latorre-Progulakis" "referencia" => array:1 [ 0 => array:2 [ "etiqueta" => "<span class="elsevierStyleSup">d</span>" "identificador" => "aff0020" ] ] ] 6 => array:3 [ "nombre" => "M." "apellidos" => "Fuentes Guajardo" "referencia" => array:1 [ 0 => array:2 [ "etiqueta" => "<span class="elsevierStyleSup">d</span>" "identificador" => "aff0020" ] ] ] 7 => array:3 [ "nombre" => "P." "apellidos" => "Martín-Escudero" "referencia" => array:1 [ 0 => array:2 [ "etiqueta" => "<span class="elsevierStyleSup">e</span>" "identificador" => "aff0025" ] ] ] ] "afiliaciones" => array:5 [ 0 => array:3 [ "entidad" => "Departamento de Ciencias de la Actividad Física y del Deporte, Facultad de Educación y Humanidades, Universidad de Tarapacá, Arica, Chile" "etiqueta" => "a" "identificador" => "aff0005" ] 1 => array:3 [ "entidad" => "Estudiante Semillero de Investigación, Departamento de Ciencias de la Actividad Física y del Deporte, Facultad de Educación y Humanidades, Universidad de Tarapacá, Arica, Chile" "etiqueta" => "b" "identificador" => "aff0010" ] 2 => array:3 [ "entidad" => "Departamento de Física, Facultad de Ciencias, Universidad de Tarapacá, Arica, Chile" "etiqueta" => "c" "identificador" => "aff0015" ] 3 => array:3 [ "entidad" => "Departamento de Tecnología Médica, Facultad de Ciencias de la Salud, Universidad de Tarapacá, Arica, Chile" "etiqueta" => "d" "identificador" => "aff0020" ] 4 => array:3 [ "entidad" => "Facultad de Medicina del Deporte, Facultad de Medicina, Universidad Complutense de Madrid, Arica, España" "etiqueta" => "e" "identificador" => "aff0025" ] ] "correspondencia" => array:1 [ 0 => array:3 [ "identificador" => "cor0005" "etiqueta" => "⁎" "correspondencia" => "Autor para correspondencia." ] ] ] ] "titulosAlternativos" => array:1 [ "en" => array:1 [ "titulo" => "Determining factors of pulse oximetry accuracy: A literature review" ] ] "resumenGrafico" => array:2 [ "original" => 0 "multimedia" => array:7 [ "identificador" => "fig0010" "etiqueta" => "Figura 2" "tipo" => "MULTIMEDIAFIGURA" "mostrarFloat" => true "mostrarDisplay" => false "figura" => array:1 [ 0 => array:4 [ "imagen" => "gr2.jpeg" "Alto" => 1788 "Ancho" => 2675 "Tamanyo" => 295878 ] ] "descripcion" => array:1 [ "es" => "<p id="spar0050" class="elsevierStyleSimplePara elsevierViewall">Localización geográfica de los artículos seleccionados.</p>" ] ] ] "textoCompleto" => "<span class="elsevierStyleSections"><span id="sec0005" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0065">Introducción</span><p id="par0005" class="elsevierStylePara elsevierViewall">La oximetría de pulso, método no invasivo para determinar la saturación de oxígeno en la sangre (SpO<span class="elsevierStyleInf">2</span>), ha adquirido gran relevancia desde el surgimiento de la COVID-19 debido a la necesidad de monitorizar esta variable clínica. Este dispositivo es práctico y proporciona resultados precisos tanto en entornos hospitalarios como domiciliarios<a class="elsevierStyleCrossRefs" href="#bib0285"><span class="elsevierStyleSup">1,2</span></a>.</p><p id="par0010" class="elsevierStylePara elsevierViewall">Su uso es extensivo en diversas especialidades médicas y en pacientes de todas las edades<a class="elsevierStyleCrossRefs" href="#bib0295"><span class="elsevierStyleSup">3,4</span></a>. Sin embargo, a pesar del incremento en el uso de la oximetría de pulso, la Administración de Alimentos y Medicamentos de los Estados Unidos, conocida como FDA, ha emitido advertencias sobre las limitaciones de los oxímetros de pulso, destacando su potencial inexactitud bajo ciertas circunstancias<a class="elsevierStyleCrossRefs" href="#bib0305"><span class="elsevierStyleSup">5,6</span></a>.</p><p id="par0015" class="elsevierStylePara elsevierViewall">En particular, la precisión de los oxímetros puede verse afectada por la altitud<a class="elsevierStyleCrossRefs" href="#bib0315"><span class="elsevierStyleSup">7,8</span></a>, el movimiento y ejercicio físico<a class="elsevierStyleCrossRef" href="#bib0325"><span class="elsevierStyleSup">9</span></a>, los diferentes tipos de hemoglobinas<a class="elsevierStyleCrossRef" href="#bib0330"><span class="elsevierStyleSup">10</span></a>, el sitio de colocación del sensor<a class="elsevierStyleCrossRef" href="#bib0335"><span class="elsevierStyleSup">11</span></a>, los colorantes de los esmaltes de uñas<a class="elsevierStyleCrossRefs" href="#bib0340"><span class="elsevierStyleSup">12,13</span></a>, la pigmentación de la piel<a class="elsevierStyleCrossRefs" href="#bib0350"><span class="elsevierStyleSup">14,15</span></a>, una arritmia cardíaca<a class="elsevierStyleCrossRef" href="#bib0360"><span class="elsevierStyleSup">16</span></a>, una baja perfusión, hipoxemia e hipoxemia postoperatoria<a class="elsevierStyleCrossRefs" href="#bib0365"><span class="elsevierStyleSup">17,18</span></a>, o el uso de mascarilla<a class="elsevierStyleCrossRef" href="#bib0375"><span class="elsevierStyleSup">19</span></a>. Por otro lado, las características técnicas del dispositivo, incluyendo los tipos de luz LED, los detectores de onda<a class="elsevierStyleCrossRef" href="#bib0380"><span class="elsevierStyleSup">20</span></a> o la medición bipolar o tetrapolar, también pueden influir en la precisión de los oxímetros<a class="elsevierStyleCrossRef" href="#bib0385"><span class="elsevierStyleSup">21</span></a>.</p><p id="par0020" class="elsevierStylePara elsevierViewall">Aunque hay numerosos estudios sobre los factores que inciden en la precisión de la oximetría de pulso, aún no se ha establecido un consenso en la literatura, debido a las distintas variables señaladas por los autores.</p><p id="par0025" class="elsevierStylePara elsevierViewall">Este estudio tiene el objetivo de examinar la literatura para identificar un acuerdo sobre las variables que afectan las mediciones de oximetría de pulso. Se buscará identificar investigaciones de la última década, utilizar el método Preferred Reporting Items for Systematic Reviews and Meta-Analyses –PRISMA– para sintetizar los factores que impactan en la precisión y determinar la distribución geográfica y temporal de los estudios que satisfacen los criterios de inclusión.</p></span><span id="sec0010" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0070">Metodología</span><p id="par0030" class="elsevierStylePara elsevierViewall">Para la búsqueda en la literatura de estudios que incluyen las variables que interfieren en los resultados de la medición de la SpO<span class="elsevierStyleInf">2</span> se realizó una recopilación de artículos científicos que estuviesen publicados en revistas indexadas, se indagó en bases de datos bibliográficos como ProQuest, ScienceDirect, Cochrane, EBSCO, Web of Science, Scopus, Epistemonikos, Google Scholar, SciELO, Dialnet y PubMed, y en sitios de Internet con acceso libre seleccionando un total de 23 artículos para la investigación (<a class="elsevierStyleCrossRef" href="#fig0005">figura 1</a> y <a class="elsevierStyleCrossRef" href="#tbl0005">tabla 1</a>).</p><elsevierMultimedia ident="fig0005"></elsevierMultimedia><elsevierMultimedia ident="tbl0005"></elsevierMultimedia><p id="par0035" class="elsevierStylePara elsevierViewall">La investigación se llevó a cabo mediante una revisión bibliográfica a través del método PRISMA<a class="elsevierStyleCrossRef" href="#bib0390"><span class="elsevierStyleSup">22</span></a>. Implementamos una búsqueda avanzada en las bases de datos citadas. Para nuestra revisión sistemática empleamos una combinación de palabras clave y descriptores específicamente seleccionados para abarcar de manera exhaustiva los términos relacionados con la oximetría de pulso y las variables que impactan las mediciones de SpO<span class="elsevierStyleInf">2</span>. Los descriptores empleados incluyeron términos primarios como «oximetría de pulso», con adaptaciones lingüísticas para permitir una búsqueda inclusiva en diversos idiomas. La estrategia de búsqueda se diseñó meticulosamente utilizando operadores booleanos para afinar los resultados. El uso del operador «AND» buscó asegurar que las publicaciones incluyeran tanto temas relacionados con la oximetría de pulso como aquellos pertinentes a la precisión y exactitud de las mediciones. Por otro lado, el operador «OR» se utilizó para conectar términos sinónimos o relacionados, ampliando así el espectro de la búsqueda para incluir un rango más amplio de estudios relevantes. Esta consulta buscó identificar literatura que no solo abordara la tecnología y las aplicaciones de la oximetría de pulso y la fotopletismografía, sino también aquella que discutiera la precisión, la exactitud y los posibles errores de medición, así como los factores que afectan estos parámetros.</p><p id="par0040" class="elsevierStylePara elsevierViewall">Se realizó un filtrado minucioso con un análisis cualitativo de los artículos seleccionados para identificar estudios que abordaran variables que influyen la medición de la SpO<span class="elsevierStyleInf">2</span>, incluyendo, pero no limitándose a, la sobrevaloración y subvaloración de la saturación, los efectos de la altitud, el movimiento, la pigmentación de la piel, el uso de tinte de henna, los niveles de perfusión, las hemoglobinopatías, la temperatura cutánea, la frecuencia cardíaca, la hipoxemia y la tecnología empleada en los oxímetros de pulso. Además, se adoptó un enfoque iterativo en nuestra búsqueda, valorando periódicamente los resultados para asegurar la pertinencia y el foco de la misma. Mantuvimos un registro detallado de la estrategia de búsqueda y los ajustes realizados durante el proceso, reiterando la búsqueda al finalizar nuestra revisión para incluir cualquier nueva literatura publicada desde el inicio de nuestra investigación. Esta búsqueda se afinó con filtros de fecha desde el año 2013 hasta el 2023, enfocándonos en artículos de revistas indexadas de acceso abierto.</p><p id="par0045" class="elsevierStylePara elsevierViewall">Procedimentalmente 3 investigadores llevaron a cabo el análisis de forma independiente, siguiendo los criterios de inclusión y exclusión predefinidos, y luego consensuaron los resultados. En casos de discrepancia, 2 investigadores intervinieron para mediar y alcanzar un acuerdo.</p><p id="par0050" class="elsevierStylePara elsevierViewall">El proceso de selección se realizó en etapas: primero, un escaneo preliminar de los títulos; seguido por una revisión detallada de los títulos y resúmenes; y, finalmente, la lectura completa de los textos que satisfacían nuestros criterios de inclusión.</p><p id="par0055" class="elsevierStylePara elsevierViewall">Los estudios seleccionados cumplían con los siguientes criterios de inclusión:</p><span id="sec0015" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0075">Criterios de inclusión</span><p id="par0060" class="elsevierStylePara elsevierViewall"><ul class="elsevierStyleList" id="lis0005"><li class="elsevierStyleListItem" id="lsti0005"><span class="elsevierStyleLabel">-</span><p id="par0065" class="elsevierStylePara elsevierViewall">Artículos con años de publicación entre 2013-2023.</p></li><li class="elsevierStyleListItem" id="lsti0010"><span class="elsevierStyleLabel">-</span><p id="par0070" class="elsevierStylePara elsevierViewall">Artículos que incluyeran variables que interfieran en el resultado de la medición de la SpO<span class="elsevierStyleInf">2</span>.</p></li><li class="elsevierStyleListItem" id="lsti0015"><span class="elsevierStyleLabel">-</span><p id="par0075" class="elsevierStylePara elsevierViewall">Investigaciones publicadas en revistas indexadas de acceso abierto.</p></li></ul></p></span><span id="sec0020" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0080">Criterios de exclusión</span><p id="par0080" class="elsevierStylePara elsevierViewall"><ul class="elsevierStyleList" id="lis0010"><li class="elsevierStyleListItem" id="lsti0020"><span class="elsevierStyleLabel">-</span><p id="par0085" class="elsevierStylePara elsevierViewall">Estudios que, a pesar de considerar la temática, no describían contenidos como las variables que intervienen en el resultado de la medición de la SpO<span class="elsevierStyleInf">2</span>.</p></li></ul></p><p id="par0090" class="elsevierStylePara elsevierViewall">Por último, descartamos estudios que, aunque relacionados con el tema, no detallaban las variables que afectan las mediciones de SpO<span class="elsevierStyleInf">2</span> o presentaban una calidad metodológica pobre de acuerdo con la Escala de Newcastle-Ottawa<a class="elsevierStyleCrossRef" href="#bib0485"><span class="elsevierStyleSup">41</span></a>. Esta escala permitió evaluar la calidad de estudios no aleatorizados en la revisión sistemática. Los estudios se juzgaron desde 3 perspectivas: la selección de los grupos de estudio; la comparabilidad de los grupos, y la determinación de la exposición o el resultado de interés para estudios de casos. Finalmente, los resultados se llevaron a una escala de evaluación de calidad de Ottawa, relacionándose su calidad con la cantidad de estrellas alcanzadas. No fueron considerados aquellos estudios que contaran con menos de 5 estrellas. Por otra parte, para el análisis de la calidad metodológica de las revisiones sistemáticas, resultó necesario utilizar una lista de comprobación de la calidad de los estudios adecuada a cada proyecto; igualmente, fue precisa la identificación de los sesgos más probables en cada uno de los diseños incluidos, lo que facilitó evaluar posteriormente la calidad de los estudios, seleccionando los que, por su puntaje, fueran clasificados como aceptables. Para ello tabulamos datos extraídos en la evaluación de calidad con el objetivo de poder clasificar con rigor los artículos de revisión sistemática.</p></span><span id="sec0025" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0085">Riesgo de sesgo en los estudios individuales</span><p id="par0095" class="elsevierStylePara elsevierViewall">El riesgo de sesgo en estudios fue valorado durante la selección de artículos usando la guía Cochrane<a class="elsevierStyleCrossRef" href="#bib0490"><span class="elsevierStyleSup">42</span></a>, considerando dominios como asignación, cegamiento, seguimientos, exclusiones y sesgos potenciales, asegurando la elección de datos con bajo riesgo de sesgo.</p></span></span><span id="sec0030" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0090">Resultados</span><p id="par0100" class="elsevierStylePara elsevierViewall">A partir del análisis mencionado se encontraron 151 artículos, de los cuales fueron seleccionados una totalidad de 23 que cumplían con los criterios de inclusión.</p><p id="par0105" class="elsevierStylePara elsevierViewall">Los artículos seleccionados procedían de las siguientes localizaciones geográficas: uno de Marruecos<a class="elsevierStyleCrossRef" href="#bib0470"><span class="elsevierStyleSup">38</span></a>, 4 de Argentina<a class="elsevierStyleCrossRefs" href="#bib0290"><span class="elsevierStyleSup">2,25,28,36</span></a>, 2 de Austria<a class="elsevierStyleCrossRefs" href="#bib0320"><span class="elsevierStyleSup">8,35</span></a>, 2 de Chile<a class="elsevierStyleCrossRefs" href="#bib0400"><span class="elsevierStyleSup">24,29</span></a>, uno de China<a class="elsevierStyleCrossRef" href="#bib0350"><span class="elsevierStyleSup">14</span></a>, 2 de Corea<a class="elsevierStyleCrossRefs" href="#bib0440"><span class="elsevierStyleSup">32,33</span></a>, 3 de Estados Unidos<a class="elsevierStyleCrossRefs" href="#bib0335"><span class="elsevierStyleSup">11,34,39</span></a>, 2 de Inglaterra<a class="elsevierStyleCrossRefs" href="#bib0435"><span class="elsevierStyleSup">31,40</span></a>, 3 de México<a class="elsevierStyleCrossRefs" href="#bib0345"><span class="elsevierStyleSup">13,23,26</span></a>, uno de Suiza<a class="elsevierStyleCrossRef" href="#bib0460"><span class="elsevierStyleSup">36</span></a>, uno de Bolivia<a class="elsevierStyleCrossRef" href="#bib0430"><span class="elsevierStyleSup">30</span></a> y otro de Colombia<a class="elsevierStyleCrossRef" href="#bib0415"><span class="elsevierStyleSup">27</span></a>. La <a class="elsevierStyleCrossRef" href="#fig0010">figura 2</a> muestra la localización geográfica de los artículos seleccionados.</p><elsevierMultimedia ident="fig0010"></elsevierMultimedia><span id="sec0035" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0095">Cronología de los artículos seleccionados</span><p id="par0110" class="elsevierStylePara elsevierViewall">Se seleccionaron artículos de los últimos 10 años, siendo los seleccionados uno de 2013<a class="elsevierStyleCrossRef" href="#bib0470"><span class="elsevierStyleSup">38</span></a>, uno de 2015<a class="elsevierStyleCrossRef" href="#bib0435"><span class="elsevierStyleSup">31</span></a>, 2 de 2016<a class="elsevierStyleCrossRefs" href="#bib0395"><span class="elsevierStyleSup">23,24</span></a>, uno de 2017<a class="elsevierStyleCrossRef" href="#bib0350"><span class="elsevierStyleSup">14</span></a>, uno de 2018<a class="elsevierStyleCrossRef" href="#bib0440"><span class="elsevierStyleSup">32</span></a>, uno de 2019<a class="elsevierStyleCrossRef" href="#bib0405"><span class="elsevierStyleSup">25</span></a>, 4 de 2020<a class="elsevierStyleCrossRefs" href="#bib0335"><span class="elsevierStyleSup">11,26,27,39</span></a>, 6 de 2021<a class="elsevierStyleCrossRefs" href="#bib0320"><span class="elsevierStyleSup">8,28,33–35,37</span></a>, 4 de 2022<a class="elsevierStyleCrossRefs" href="#bib0345"><span class="elsevierStyleSup">13,29,36,40</span></a> y 2 de 2023<a class="elsevierStyleCrossRefs" href="#bib0290"><span class="elsevierStyleSup">2,30</span></a>. El máximo número de publicaciones fue en 2021, posiblemente debido al incremento del uso de la oximetría de pulso durante la pandemia de COVID-19 y la atención a la precisión de las mediciones de SpO<span class="elsevierStyleInf">2</span>.</p></span><span id="sec0040" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0100">Resultados del análisis de la literatura</span><p id="par0115" class="elsevierStylePara elsevierViewall">Como principales resultados (<a class="elsevierStyleCrossRef" href="#tbl0010">tabla 2</a>), se encontraron 5 artículos en los que se resaltó la tecnología como uno de los factores más influyentes en la medición de la SpO<span class="elsevierStyleInf">2</span>. Un estudio comprobó<a class="elsevierStyleCrossRef" href="#bib0400"><span class="elsevierStyleSup">24</span></a> que los 2 principales factores de error hallados fueron los artefactos por movimiento y la hipoperfusión. Destacaba que cuando el paciente se mueve aparece una «onda de pulso venosa», no siendo capaz el oxímetro de diferenciarla de la onda de pulso arterial por ser la saturación de la sangre venosa menor que la arterial, calculándose un registro menor del real. En la hipoperfusión, la baja intensidad de la señal generó un registro menor del real.</p><elsevierMultimedia ident="tbl0010"></elsevierMultimedia><p id="par0120" class="elsevierStylePara elsevierViewall">En otro estudio<a class="elsevierStyleCrossRef" href="#bib0290"><span class="elsevierStyleSup">2</span></a> los autores diseñaron un oxímetro de pulso de bajo costo, con la finalidad de contribuir a una mayor asequibilidad de estos equipos; para ello, incluyeron, en su construcción, placa MAX30102, placa Arduino UNO y un visor OLED SSD1306. Luego compararon este producto con métodos estándar, concluyendo que la medición de la SpO<span class="elsevierStyleInf">2</span> fue similar, recomendándose probar el sistema en su totalidad antes de ser lanzado al mercado.</p><p id="par0125" class="elsevierStylePara elsevierViewall">Asimismo, Santos et al.<a class="elsevierStyleCrossRef" href="#bib0410"><span class="elsevierStyleSup">26</span></a> reflexionaron sobre elementos tecnológicos que pueden tributar en la calidad de la medición de la saturación de 3 oxímetros (Onyx II, Hergom y SB220), destacando cierto grado de variabilidad. Los autores señalaron que el SB220 presenta una amplia variabilidad, mientras que el Onyx II y el Hergom mantuvieron la exactitud y la precisión. Enfatizaron que las variabilidades que más influyen en la medición estaban relacionadas con problemas de calidad en la obtención de la señal fotopletismográfica, en el funcionamiento del instrumento, en la baja calidad de la fabricación, así como en el lugar de la medición. La desaturación rápida, la hipotensión, la hipotermia, la dishemoglobinemia, la hipoperfusión y la hipoxemia aguda y crónica fueron factores que también influyeron en el resultado.</p><p id="par0130" class="elsevierStylePara elsevierViewall">Por otro lado, un artículo<a class="elsevierStyleCrossRef" href="#bib0395"><span class="elsevierStyleSup">23</span></a> consideró la utilización de oxímetros tetrapolares, empleando la técnica de impedanciometría en su configuración tetrapolar para obtener de forma más amplia la señal pletismográfica. En este punto, 2 artículos<a class="elsevierStyleCrossRefs" href="#bib0405"><span class="elsevierStyleSup">25,27</span></a> subrayaron la importancia de considerar tecnologías como la espectroscopia en la cercanía del espectro infrarrojo (NIRS) como una herramienta necesaria a emplear en la atención clínica de neonatos, señalando que esta tecnología permite realizar la medición del consumo de oxígeno en tiempo real y de forma no invasiva en diversos tejidos somáticos, como el cerebro, el intestino y los riñones.</p><p id="par0135" class="elsevierStylePara elsevierViewall">Se consideraron 4 artículos<a class="elsevierStyleCrossRefs" href="#bib0335"><span class="elsevierStyleSup">11,28,37,40</span></a> relacionados con la tecnología de los oxímetros utilizados en pacientes con insuficiencia respiratoria por COVID-19 y diferentes enfermedades pulmonares; en ellos se resaltó la necesidad de mejorar la tecnología para la evaluación de la SpO<span class="elsevierStyleInf">2</span>, acentuando que si bien se han creado saturómetros con tecnología avanzada, aún existen falencias en los mismos, quedando pendientes algunas variables que pueden afectar la exactitud de la monitorización, señalando: hablar, reír, realización de actividad física, falta del flujo pulsátil, precisión en pacientes con hipoxemia, hiperbilirrubinemia, dishemoglobinemias, movimiento excesivo, interferencia de la luz ambiental, esmalte de uñas, mayor pigmentación de la piel, individuos con un índice de masa corporal elevado, entre otros.</p><p id="par0140" class="elsevierStylePara elsevierViewall">Otros 2 artículos<a class="elsevierStyleCrossRefs" href="#bib0320"><span class="elsevierStyleSup">8,30</span></a> señalaron que existen resultados falseados por las interferencias ambientales, como la altitud. En este caso el organismo humano está expuesto a una disponibilidad reducida de oxígeno (hipoxia) debido a la disminución de la presión barométrica y la presión parcial de oxígeno asociada; como consecuencia, se inician varias respuestas fisiológicas para contrarrestar la hipoxia y el riesgo asociado de enfermarse. En el estudio, los autores hacen referencia a Rahn y Otis<a class="elsevierStyleCrossRef" href="#bib0495"><span class="elsevierStyleSup">43</span></a> y a Lenfant y Sullivan<a class="elsevierStyleCrossRef" href="#bib0500"><span class="elsevierStyleSup">44</span></a>, quienes plantearon que durante los primeros días a gran altura se produce, en comparación con la exposición aguda, un aumento progresivo de la ventilación en reposo (aclimatación ventilatoria) acompañada de un incremento de la presión parcial de oxígeno arterial y una disminución de la presión parcial alveolar de dióxido de carbono. Del mismo modo, destacaron lo planteado por Sato et al.<a class="elsevierStyleCrossRef" href="#bib0505"><span class="elsevierStyleSup">45</span></a> y Bender et al.<a class="elsevierStyleCrossRef" href="#bib0510"><span class="elsevierStyleSup">46</span></a>, los que plasmaron que la aclimatación ventilatoria se caracteriza por una disminución inicial y un aumento posterior de los valores de presión parcial de oxígeno en reposo, alcanzando un máximo en un período de entre 4 y 8 días, al menos a una altitud de hasta 4.300<span class="elsevierStyleHsp" style=""></span>m s. n. m. A diferencia de las estancias pasivas en altitud, las actividades físicas en altitud pueden retrasar la recuperación de la presión parcial de oxígeno.</p><p id="par0145" class="elsevierStylePara elsevierViewall">Un artículo hace referencia a la interferencia lumínica<a class="elsevierStyleCrossRef" href="#bib0445"><span class="elsevierStyleSup">33</span></a>; en él se destaca que la luz ambiental interfiere en las longitudes de onda verde (500 a 550<span class="elsevierStyleHsp" style=""></span>nm) y roja (645 a 700<span class="elsevierStyleHsp" style=""></span>nm). Esto se debe a que los coeficientes de absorción de la hemoglobina oxigenada y desoxigenada son opuestos en las longitudes de onda azul (380 a 500<span class="elsevierStyleHsp" style=""></span>nm) y roja. En el caso de la luz verde, la hemoglobina oxigenada absorbe la luz verde con más fuerza que la hemoglobina desoxigenada. Por otro lado, en el caso de la luz roja, la hemoglobina desoxigenada absorbe la luz roja con más fuerza que la hemoglobina oxigenada.</p><p id="par0150" class="elsevierStylePara elsevierViewall">De la misma manera, se hallaron 7 artículos<a class="elsevierStyleCrossRefs" href="#bib0350"><span class="elsevierStyleSup">14,29,32,36,39</span></a> en los que se evaluó la influencia de la pigmentación de la piel; estos autores destacaron que los pacientes de pigmentación oscura mostraban 3 veces más frecuencia de hipoxemia oculta, la cual puede no ser detectada por oximetría de pulso, ya que los oxímetros de pulso cuentan con limitaciones al evaluar sujetos de piel oscura, especialmente en condiciones de baja saturación o hipoxemia. También hicieron referencia a la interferencia de la melanina con la calidad de la señal reflejada, es decir, el aumento de la pigmentación puede absorber la luz emitida por los equipos e interferir en la detección y medida de la hemoglobina oxigenada.</p><p id="par0155" class="elsevierStylePara elsevierViewall">En un artículo<a class="elsevierStyleCrossRef" href="#bib0345"><span class="elsevierStyleSup">13</span></a> se menciona que el esmalte de uñas ecológico disminuyó los valores de saturación, los cuales se incrementaron luego del retiro del esmalte de uñas. Los autores destacaron que el esmalte de uña ecológico puede absorber la luz emitida por los aparatos e interferir en la detección y medida de la hemoglobina oxigenada y, además, puede interferir con la capacidad del oxímetro de pulso para interpretar la SpO<span class="elsevierStyleInf">2</span>.</p><p id="par0160" class="elsevierStylePara elsevierViewall">En otro artículo<a class="elsevierStyleCrossRef" href="#bib0470"><span class="elsevierStyleSup">38</span></a> mencionaron que los tintes de la piel y la henna negra interfieren en las mediciones. Los autores destacaron que la henna provocaba una coloración negra de la piel durante la fase inicial, absorbiendo todas las longitudes de onda de la luz visible, dejando pasar únicamente los rayos infrarrojos; esto da como resultado que no se pueda monitorizar la SpO<span class="elsevierStyleInf">2</span> usando oxímetros de pulso, ya que sus resultados serían inexactos.</p><p id="par0165" class="elsevierStylePara elsevierViewall">Se analizó un artículo<a class="elsevierStyleCrossRef" href="#bib0435"><span class="elsevierStyleSup">31</span></a> en el que se hace referencia a que el empleo de NIRS ofrece más parámetros para evaluar el suministro y el consumo de oxígeno en los tejidos profundos, en cambio, el oxímetro de pulso solo evalúa el estado del suministro de oxígeno. En su estudio enfatizaron que la evaluación en el dedo con NIRS y la presión parcial de oxígeno del antebrazo, en pacientes con oclusión venosa, no mostró ninguna caída significativa con respecto al valor inicial. Sin embargo, durante la oclusión total, todos los índices indicaron cambios en la perfusión sanguínea. Los cambios de desoxihemoglobina, oxihemoglobina y hemoglobina total, estimados por fotopletismografía, presentaron una alta correlación con los mismos parámetros obtenidos por el empleo de NIRS en el transcurso de oclusión venosa y durante la oclusión total. El sistema demostró la capacidad de extraer información valiosa de las señales fotopletismografía para una evaluación más amplia de la perfusión sanguínea de los tejidos.</p><p id="par0170" class="elsevierStylePara elsevierViewall">En un artículo<a class="elsevierStyleCrossRef" href="#bib0455"><span class="elsevierStyleSup">35</span></a> se comparó la monitorización no invasiva de SpO<span class="elsevierStyleInf">2</span> mediante oximetría de pulso con mediciones de SpO<span class="elsevierStyleInf">2</span> a partir de muestras de sangre en bebés prematuros y a término, e incluyeron mediciones de hemoglobina fetal. En sus informes, la mayoría encontró un sesgo de saturación parcial de oxígeno, pero aún no está claro si esto puede explicarse por las elevadas fracciones de hemoglobina fetal únicamente en la sangre neonatal. Según las características fisiológicas de la disociación de la oxihemoglobina fetal, podría haber una influencia de la hemoglobina fetal en las lecturas de SpO<span class="elsevierStyleInf">2</span>, lo que resultaría en una sobreestimación de la SpO<span class="elsevierStyleInf">2</span> para los rangos de saturación más bajos.</p></span></span><span id="sec0045" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0105">Discusión</span><p id="par0175" class="elsevierStylePara elsevierViewall">En esta revisión sistemática se identificaron factores clave que modifican la precisión en la oximetría de pulso. La tecnología de los dispositivos influye notablemente en la calidad de la medición, siendo los oxímetros tetrapolares destacados por su fiabilidad<a class="elsevierStyleCrossRefs" href="#bib0290"><span class="elsevierStyleSup">2,24,26</span></a>. Estos hallazgos concuerdan con los de Barea<a class="elsevierStyleCrossRef" href="#bib0515"><span class="elsevierStyleSup">47</span></a> y Hernández<a class="elsevierStyleCrossRef" href="#bib0520"><span class="elsevierStyleSup">48</span></a>, quienes indican que la calidad de la señal y la superficie de contacto de los electrodos son determinantes.</p><p id="par0180" class="elsevierStylePara elsevierViewall">Además, se enfatiza la necesidad de mejorar la tecnología en la evaluación de la SpO<span class="elsevierStyleInf">2</span> en contextos clínicos, como la insuficiencia respiratoria y otras enfermedades pulmonares, particularmente en la era COVID-19<a class="elsevierStyleCrossRefs" href="#bib0335"><span class="elsevierStyleSup">11,28,37,40</span></a>. Berna y Mendoza<a class="elsevierStyleCrossRef" href="#bib0370"><span class="elsevierStyleSup">18</span></a> y Jubran<a class="elsevierStyleCrossRef" href="#bib0525"><span class="elsevierStyleSup">49</span></a> junto con Hess<a class="elsevierStyleCrossRef" href="#bib0530"><span class="elsevierStyleSup">50</span></a> también resaltan diferencias en la precisión según la SpO<span class="elsevierStyleInf">2</span> medida por gasometría.</p><p id="par0185" class="elsevierStylePara elsevierViewall">La altitud es otro factor ambiental crítico que afecta las mediciones de SpO<span class="elsevierStyleInf">2</span> debido a la disminución de la presión barométrica y la presión parcial de oxígeno, como se observa en los estudios de Novillo y Mata<a class="elsevierStyleCrossRef" href="#bib0535"><span class="elsevierStyleSup">51</span></a> y los 2 artículos seleccionados<a class="elsevierStyleCrossRefs" href="#bib0320"><span class="elsevierStyleSup">8,30</span></a>.</p><p id="par0190" class="elsevierStylePara elsevierViewall">La interferencia lumínica es reconocida como un elemento perturbador de las mediciones de saturación, según un artículo<a class="elsevierStyleCrossRef" href="#bib0445"><span class="elsevierStyleSup">33</span></a> y corroborado por Mejía Salas y Mejía Suárez<a class="elsevierStyleCrossRef" href="#bib0365"><span class="elsevierStyleSup">17</span></a>, quienes recomiendan cubrir el sensor para evitar este problema.</p><p id="par0195" class="elsevierStylePara elsevierViewall">En cuanto a la pigmentación de la piel, varios estudios<a class="elsevierStyleCrossRefs" href="#bib0350"><span class="elsevierStyleSup">14,29,32,34,36,39</span></a> apoyan la evidencia de la FDA<a class="elsevierStyleCrossRef" href="#bib0310"><span class="elsevierStyleSup">6</span></a> y Bickler et al.<a class="elsevierStyleCrossRef" href="#bib0540"><span class="elsevierStyleSup">52</span></a>, quienes afirman que la pigmentación oscura puede resultar en lecturas sobreestimadas de la SpO<span class="elsevierStyleInf">2</span>.</p><p id="par0200" class="elsevierStylePara elsevierViewall">La presencia de esmalte de uñas ecológico y tintes como la henna negra también se identificaron como factores que afectan la precisión del oxímetro<a class="elsevierStyleCrossRefs" href="#bib0345"><span class="elsevierStyleSup">13,38</span></a>. Estos resultados difieren de los reportados por Çiçek et al.<a class="elsevierStyleCrossRef" href="#bib0545"><span class="elsevierStyleSup">53</span></a>, quienes no encontraron diferencias significativas con diferentes colores de esmalte.</p><p id="par0205" class="elsevierStylePara elsevierViewall">Por último, se examinó la utilidad de la NIRS como técnica complementaria para evaluar la saturación y el consumo de oxígeno en tejidos profundos<a class="elsevierStyleCrossRef" href="#bib0435"><span class="elsevierStyleSup">31</span></a>. Este enfoque coincide con la perspectiva de Casasola<a class="elsevierStyleCrossRef" href="#bib0550"><span class="elsevierStyleSup">54</span></a>, quien destaca la influencia de la melanina en los resultados de la medición.</p><p id="par0210" class="elsevierStylePara elsevierViewall">Finalmente, Pritišanac et al.<a class="elsevierStyleCrossRef" href="#bib0455"><span class="elsevierStyleSup">35</span></a> abordan la posible influencia de la hemoglobina fetal en las lecturas de saturación en neonatos, un aspecto que no encuentra consenso con Ortega<a class="elsevierStyleCrossRef" href="#bib0555"><span class="elsevierStyleSup">55</span></a>, quien plantea que la hemoglobina anormal puede no afectar la oximetría de pulso, pero sí la oxigenación tisular.</p><p id="par0215" class="elsevierStylePara elsevierViewall">Como resumen, la <a class="elsevierStyleCrossRef" href="#tbl0015">tabla 3</a> muestra las principales limitaciones de los oxímetros de pulso divididos en 3 categorías principales: factores que causan subestimación de la SpO<span class="elsevierStyleInf">2</span>, factores que conducen a una sobreestimación de la SpO<span class="elsevierStyleInf">2</span> y factores que distorsionan la señal del fotopletismograma<a class="elsevierStyleCrossRef" href="#bib0560"><span class="elsevierStyleSup">56</span></a>.</p><elsevierMultimedia ident="tbl0015"></elsevierMultimedia><p id="par0220" class="elsevierStylePara elsevierViewall">La medición de la SpO<span class="elsevierStyleInf">2</span> puede ser subestimada por factores como la perfusión periférica deficiente, el grosor y la baja temperatura de la piel, el uso de esmalte de uñas o tatuajes, la anemia hipoxémica, la baja perfusión, el entrenamiento en altitudes elevadas, el uso de bloqueadores de nervios y la presencia de tintes intravenosos. Por otro lado, la sobreestimación de SpO<span class="elsevierStyleInf">2</span> puede ser causada por la pigmentación de la piel, hemoglobinopatías, alta temperatura cutánea o el hábito de fumar. Además, la señal del fotopletismograma puede distorsionarse por artefactos de movimiento, contaminación lumínica, frecuencias de bombeo elevadas, interferencia electromagnética, colocación incorrecta de la sonda, y ritmos cardíacos irregulares o arritmias.</p><p id="par0225" class="elsevierStylePara elsevierViewall">Estos hallazgos resaltan la necesidad de considerar una gama de variables fisiológicas, ambientales y técnicas al interpretar los resultados de la oximetría de pulso para garantizar la precisión y la fiabilidad en la evaluación clínica.</p><p id="par0230" class="elsevierStylePara elsevierViewall">En nuestro estudio consideramos como fortaleza una metodología rigurosa con uso de criterios de inclusión y exclusión estrictos, junto con una revisión sistemática de bases de datos actualizadas hasta diciembre de 2023, asegurando la relevancia y calidad de los estudios incluidos. La revisión abarcó una diversa gama de bases de datos y sitios de Internet con acceso libre, seleccionando un total de 23 artículos relevantes, lo que indica una exhaustiva cobertura del tema. Por otra parte, los resultados subrayan la importancia de considerar múltiples variables al interpretar las mediciones de oximetría de pulso, lo que tiene implicaciones importantes para la práctica clínica y la investigación futura. Sin embargo, señalamos como posibles debilidades en nuestra investigación la variabilidad y heterogeneidad en los diseños de estudios incluidos, lo que puede limitar la consistencia y precisión de la evidencia. Por otra parte, aunque se resalta la importancia de la tecnología en la precisión de las mediciones, el estudio reconoce que aún existen falencias en las tecnologías actuales de los oxímetros de pulso. Igualmente, se evidenció una carencia de investigaciones que consideren la influencia de factores fisiológicos y ambientales, los cuales no se encuentran ampliamente explorados, por lo que el estudio apunta a que aún es necesario explorar más a fondo y en mayor detalle las causas que pueden alterar los resultados de la SpO<span class="elsevierStyleInf">2</span>.</p></span><span id="sec0050" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0110">Conclusión</span><p id="par0235" class="elsevierStylePara elsevierViewall">Los hallazgos de este estudio indican que la tecnología implementada en los oxímetros es un factor crítico para la precisión de las mediciones, siendo los dispositivos tetrapolares superiores debido a su mayor exactitud. En contraste, tecnologías menos avanzadas tienden a ser menos precisas, especialmente en pacientes con compromisos respiratorios, como aquellos afectados por COVID-19.</p><p id="par0240" class="elsevierStylePara elsevierViewall">Entre los factores ambientales, la altitud se destacó por causar bajos niveles de SpO<span class="elsevierStyleInf">2</span> debido a la hipoxia ambiental. Paralelamente, se observó que la luz ambiental puede alterar significativamente la medición al interferir con las longitudes de onda verde y roja, donde la hemoglobina desoxigenada absorbe más luz roja que la oxigenada, afectando así la fiabilidad de los resultados de oximetría.</p><p id="par0245" class="elsevierStylePara elsevierViewall">Otros elementos identificados incluyen el sesgo debido al color de la piel y factores como el esmalte de uñas, tintes de la piel y henna negra. Además, se reconoce la relevancia de la NIRS y la necesidad de considerar las mediciones de la SpO<span class="elsevierStyleInf">2</span> en sangre para neonatos y bebés a término como variables que pueden influir en los resultados de la oximetría de pulso.</p><p id="par0250" class="elsevierStylePara elsevierViewall">Estos hallazgos resaltan la necesidad de considerar una gama de variables fisiológicas, ambientales y técnicas al interpretar los resultados de la oximetría de pulso para garantizar la precisión y la fiabilidad en la evaluación clínica.</p><p id="par0255" class="elsevierStylePara elsevierViewall">Dada la complejidad de los factores que afectan la precisión de la oximetría de pulso, se recomienda ampliar la investigación para incluir un mayor número de referencias literarias. Es crucial considerar una diversidad de grupos etarios y géneros y examinar con más detalle las causas que pueden alterar los resultados de la SpO<span class="elsevierStyleInf">2</span>. Estos esfuerzos pueden contribuir a mejorar la exactitud del diagnóstico y el seguimiento de los pacientes, optimizando así la atención clínica.</p></span><span id="sec0055" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0115">Financiación</span><p id="par0260" class="elsevierStylePara elsevierViewall">Esta revisión fue financiada por el proyecto número SA2210178.</p></span><span id="sec0060" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0120">Conflicto de intereses</span><p id="par0265" class="elsevierStylePara elsevierViewall">Los autores niegan tener conflicto de interés de cualquier índole.</p></span></span>" "textoCompletoSecciones" => array:1 [ "secciones" => array:13 [ 0 => array:3 [ "identificador" => "xres2141460" "titulo" => "Resumen" "secciones" => array:4 [ 0 => array:2 [ "identificador" => "abst0005" "titulo" => "Objetivo" ] 1 => array:2 [ "identificador" => "abst0010" "titulo" => "Métodos" ] 2 => array:2 [ "identificador" => "abst0015" "titulo" => "Resultados" ] 3 => array:2 [ "identificador" => "abst0020" "titulo" => "Conclusiones" ] ] ] 1 => array:2 [ "identificador" => "xpalclavsec1818027" "titulo" => "Palabras clave" ] 2 => array:3 [ "identificador" => "xres2141461" "titulo" => "Abstract" "secciones" => array:4 [ 0 => array:2 [ "identificador" => "abst0025" "titulo" => "Objective" ] 1 => array:2 [ "identificador" => "abst0030" "titulo" => "Methods" ] 2 => array:2 [ "identificador" => "abst0035" "titulo" => "Results" ] 3 => array:2 [ "identificador" => "abst0040" "titulo" => "Conclusions" ] ] ] 3 => array:2 [ "identificador" => "xpalclavsec1818028" "titulo" => "Keywords" ] 4 => array:2 [ "identificador" => "sec0005" "titulo" => "Introducción" ] 5 => array:3 [ "identificador" => "sec0010" "titulo" => "Metodología" "secciones" => array:3 [ 0 => array:2 [ "identificador" => "sec0015" "titulo" => "Criterios de inclusión" ] 1 => array:2 [ "identificador" => "sec0020" "titulo" => "Criterios de exclusión" ] 2 => array:2 [ "identificador" => "sec0025" "titulo" => "Riesgo de sesgo en los estudios individuales" ] ] ] 6 => array:3 [ "identificador" => "sec0030" "titulo" => "Resultados" "secciones" => array:2 [ 0 => array:2 [ "identificador" => "sec0035" "titulo" => "Cronología de los artículos seleccionados" ] 1 => array:2 [ "identificador" => "sec0040" "titulo" => "Resultados del análisis de la literatura" ] ] ] 7 => array:2 [ "identificador" => "sec0045" "titulo" => "Discusión" ] 8 => array:2 [ "identificador" => "sec0050" "titulo" => "Conclusión" ] 9 => array:2 [ "identificador" => "sec0055" "titulo" => "Financiación" ] 10 => array:2 [ "identificador" => "sec0060" "titulo" => "Conflicto de intereses" ] 11 => array:2 [ "identificador" => "xack743267" "titulo" => "Agradecimientos" ] 12 => array:1 [ "titulo" => "Bibliografía" ] ] ] "pdfFichero" => "main.pdf" "tienePdf" => true "fechaRecibido" => "2024-01-18" "fechaAceptado" => "2024-03-07" "PalabrasClave" => array:2 [ "es" => array:1 [ 0 => array:4 [ "clase" => "keyword" "titulo" => "Palabras clave" "identificador" => "xpalclavsec1818027" "palabras" => array:5 [ 0 => "Saturación de oxígeno" 1 => "Pigmentación de la piel" 2 => "Exactitud" 3 => "Precisión" 4 => "Fotopletismografía" ] ] ] "en" => array:1 [ 0 => array:4 [ "clase" => "keyword" "titulo" => "Keywords" "identificador" => "xpalclavsec1818028" "palabras" => array:5 [ 0 => "Oxygen saturation" 1 => "Skin pigmentation" 2 => "Accuracy" 3 => "Precision" 4 => "Photoplethysmography" ] ] ] ] "tieneResumen" => true "resumen" => array:2 [ "es" => array:3 [ "titulo" => "Resumen" "resumen" => "<span id="abst0005" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0010">Objetivo</span><p id="spar0005" class="elsevierStyleSimplePara elsevierViewall">Identificar y alcanzar un consenso sobre las variables que afectan la medición de la saturación de oxígeno (SpO<span class="elsevierStyleInf">2</span>) mediante la oximetría de pulso.</p></span> <span id="abst0010" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0015">Métodos</span><p id="spar0010" class="elsevierStyleSimplePara elsevierViewall">Aplicamos criterios de inclusión y exclusión para seleccionar estudios relevantes en bases de datos como EBSCO y PubMed. Las estrategias de búsqueda, realizadas hasta diciembre de 2023, se centraron en publicaciones que abordaran la tecnología de los oxímetros de pulso y variables que influyen en su precisión. Evaluamos el riesgo de sesgo de los estudios incluidos y utilizamos métodos estandarizados para la síntesis de resultados.</p></span> <span id="abst0015" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0020">Resultados</span><p id="spar0015" class="elsevierStyleSimplePara elsevierViewall">Se incluyeron 23 estudios. La síntesis de los resultados destacó que los equipos con tecnología tetrapolar mostraron mayor precisión en las mediciones de SpO<span class="elsevierStyleInf">2</span>. Una mayor pigmentación de la piel, hemoglobinopatías y altas temperaturas de la piel pueden llevar a una sobreestimación de la SpO<span class="elsevierStyleInf">2</span>, mientras que factores como baja perfusión, temperatura de la piel fría, esmalte de uñas o tatuajes, hipoxemia, anemia y entrenamiento a gran altitud pueden subestimarla. Por otra parte, artefactos de movimiento, contaminación lumínica, frecuencia<span class="elsevierStyleHsp" style=""></span>><span class="elsevierStyleHsp" style=""></span>150<span class="elsevierStyleHsp" style=""></span>lpm, interferencia electromagnética y el lugar de colocación del sensor pueden provocar distorsión de la señal de fotopletismografía.</p></span> <span id="abst0020" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0025">Conclusiones</span><p id="spar0020" class="elsevierStyleSimplePara elsevierViewall">La síntesis de los resultados destacó que la pigmentación de la piel y las interferencias luminosas pueden llevar a una sobreestimación de la SpO<span class="elsevierStyleInf">2</span>, mientras que otros factores, como la baja perfusión y la altitud, tienden a subestimarla. Los estudios presentaron variabilidad y heterogeneidad en sus diseños, evidenciando limitaciones en la consistencia y precisión de la evidencia. A pesar de estas limitaciones, los resultados subrayan la importancia de considerar múltiples variables al interpretar las mediciones de la oximetría de pulso para garantizar su confiabilidad. Los hallazgos tienen implicaciones significativas para la práctica clínica y la investigación futura.</p></span>" "secciones" => array:4 [ 0 => array:2 [ "identificador" => "abst0005" "titulo" => "Objetivo" ] 1 => array:2 [ "identificador" => "abst0010" "titulo" => "Métodos" ] 2 => array:2 [ "identificador" => "abst0015" "titulo" => "Resultados" ] 3 => array:2 [ "identificador" => "abst0020" "titulo" => "Conclusiones" ] ] ] "en" => array:3 [ "titulo" => "Abstract" "resumen" => "<span id="abst0025" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0035">Objective</span><p id="spar0025" class="elsevierStyleSimplePara elsevierViewall">Identify and reach consensus on the variables that affect the measurement of oxygen saturation (SpO<span class="elsevierStyleInf">2</span>) using pulse oximetry.</p></span> <span id="abst0030" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0040">Methods</span><p id="spar0030" class="elsevierStyleSimplePara elsevierViewall">We applied inclusion and exclusion criteria to select relevant studies in databases such as EBSCO and PubMed. The search strategies, carried out until December 2023, focused on publications that addressed the technology of pulse oximeters and variables that influence their accuracy. We assessed the risk of bias of the included studies and used standardized methods for synthesis of results.</p></span> <span id="abst0035" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0045">Results</span><p id="spar0035" class="elsevierStyleSimplePara elsevierViewall">Twenty-three studies were included. The synthesis of the results highlighted that equipment with tetrapolar technology showed greater precision in SpO<span class="elsevierStyleInf">2</span> measurements. Increased skin pigmentation, hemoglobinopathies and high skin temperatures can lead to an overestimation of SpO<span class="elsevierStyleInf">2</span>, while factors such as low perfusion, cold skin temperature, nail polish or tattoos, hypoxemia, anemia and high altitude training, they may underestimate it. On the other hand, motion artifacts, light pollution, frequency<span class="elsevierStyleHsp" style=""></span>><span class="elsevierStyleHsp" style=""></span>150<span class="elsevierStyleHsp" style=""></span>bpm, electromagnetic interference and location of the sensor can cause distortion of the photoplethymography signal.</p></span> <span id="abst0040" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0050">Conclusions</span><p id="spar0040" class="elsevierStyleSimplePara elsevierViewall">The synthesis of the results highlighted that skin pigmentation and light interference can lead to an overestimation of SpO<span class="elsevierStyleInf">2</span>, while other factors such as low perfusion and altitude tend to underestimate it. The studies presented variability and heterogeneity in their designs, evidencing limitations in the consistency and precision of the evidence. Despite these limitations, the results underscore the importance of considering multiple variables when interpreting pulse oximetry measurements to ensure their reliability. The findings have significant implications for clinical practice and future research.</p></span>" "secciones" => array:4 [ 0 => array:2 [ "identificador" => "abst0025" "titulo" => "Objective" ] 1 => array:2 [ "identificador" => "abst0030" "titulo" => "Methods" ] 2 => array:2 [ "identificador" => "abst0035" "titulo" => "Results" ] 3 => array:2 [ "identificador" => "abst0040" "titulo" => "Conclusions" ] ] ] ] "multimedia" => array:5 [ 0 => array:7 [ "identificador" => "fig0005" "etiqueta" => "Figura 1" "tipo" => "MULTIMEDIAFIGURA" "mostrarFloat" => true "mostrarDisplay" => false "figura" => array:1 [ 0 => array:4 [ "imagen" => "gr1.jpeg" "Alto" => 1343 "Ancho" => 1667 "Tamanyo" => 152618 ] ] "descripcion" => array:1 [ "es" => "<p id="spar0045" class="elsevierStyleSimplePara elsevierViewall">Diagrama de flujo consistente con la declaración de elementos de informe preferidos para revisiones sistemáticas y metaanálisis (PRISMA).</p>" ] ] 1 => array:7 [ "identificador" => "fig0010" "etiqueta" => "Figura 2" "tipo" => "MULTIMEDIAFIGURA" "mostrarFloat" => true "mostrarDisplay" => false "figura" => array:1 [ 0 => array:4 [ "imagen" => "gr2.jpeg" "Alto" => 1788 "Ancho" => 2675 "Tamanyo" => 295878 ] ] "descripcion" => array:1 [ "es" => "<p id="spar0050" class="elsevierStyleSimplePara elsevierViewall">Localización geográfica de los artículos seleccionados.</p>" ] ] 2 => array:8 [ "identificador" => "tbl0005" "etiqueta" => "Tabla 1" "tipo" => "MULTIMEDIATABLA" "mostrarFloat" => true "mostrarDisplay" => false "detalles" => array:1 [ 0 => array:3 [ "identificador" => "at1" "detalle" => "Tabla " "rol" => "short" ] ] "tabla" => array:1 [ "tablatextoimagen" => array:1 [ 0 => array:2 [ "tabla" => array:1 [ 0 => """ <table border="0" frame="\n \t\t\t\t\tvoid\n \t\t\t\t" class=""><thead title="thead"><tr title="table-row"><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Base de Datos \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Nombre del artículo \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Autores \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Codificación para estudios de casos y controlesselección \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Análisis calidad metodológica revisión sistemática \t\t\t\t\t\t\n \t\t\t\t\t\t</th></tr></thead><tbody title="tbody"><tr title="table-row"><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " rowspan="10" align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Google Scholar</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Pletismografía de impedancia basada en electrodos de Polipirrol para el sensado de la frecuencia cardiaca \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Hernández & Suaste, (2016)<a class="elsevierStyleCrossRef" href="#bib0395"><span class="elsevierStyleSup">23</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">8 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Oximetría de pulso: ¿cuál es el equipo ideal? \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Perillán, (2016)<a class="elsevierStyleCrossRef" href="#bib0400"><span class="elsevierStyleSup">24</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Evaluación cualitativa- aceptable(Revisión bibliográfica) \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Aplicación de nuevas tecnologías en el cuidado neonatal: monitorización de la saturación regional de oxígeno \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Arnaudo, (2019)<a class="elsevierStyleCrossRef" href="#bib0405"><span class="elsevierStyleSup">25</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Evaluación cualitativa- aceptable(Revisión bibliográfica) \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Diferencias en la saturación parcial de oxígeno obtenida de tres oxímetros de pulso \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Santos, et al. (2020)<a class="elsevierStyleCrossRef" href="#bib0410"><span class="elsevierStyleSup">26</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">8 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">¿La espectroscopia de infrarrojo cercano (NIRS) podría ser útil en el monitoreo hemodinámico del paciente critico pediátrico? \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Puentes, et al. (2020)<a class="elsevierStyleCrossRef" href="#bib0415"><span class="elsevierStyleSup">27</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Evaluación cualitativa-aceptable(Revisión bibliográfica) \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Estudio retrospectivo para describir la concordancia entre dos oxímetros de pulso \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Bosio, et al. (2021)<a class="elsevierStyleCrossRef" href="#bib0420"><span class="elsevierStyleSup">28</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">9 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Esmaltes de uñas ecológicos y medición de la saturación periférica de oxígeno en pacientes quirúrgicos. Presentación de casos \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Díaz, et al. (2022)<a class="elsevierStyleCrossRef" href="#bib0345"><span class="elsevierStyleSup">13</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">9 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Skin pigmentation effect on the accuracy of pulse oximetry: a systematic review \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Cabanas, et al. (2022)<a class="elsevierStyleCrossRef" href="#bib0425"><span class="elsevierStyleSup">29</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Aceptable \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Diseño de un oxímetro de pulso \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Barrera, et al. (2023)<a class="elsevierStyleCrossRef" href="#bib0290"><span class="elsevierStyleSup">2</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">5 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Variaciones fenotípicas relacionadas a respuestas hipóxicas entre andinos a distintas altitudes \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Luna, et al. (2023)<a class="elsevierStyleCrossRef" href="#bib0430"><span class="elsevierStyleSup">30</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">8 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " rowspan="7" align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">CrossRef</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Reflectance photoplethysmography as noninvasive monitoring of tissue blood perfusion \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Abay & Kyriacou, (2015)<a class="elsevierStyleCrossRef" href="#bib0435"><span class="elsevierStyleSup">31</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">8 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">The effect of optical crosstalk on accuracy of reflectance-type pulse oximeter for mobile healthcare \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Baek, et al. (2018)<a class="elsevierStyleCrossRef" href="#bib0440"><span class="elsevierStyleSup">32</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">6 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Non-contact oxygen saturation measurement using YCgCr color space with an RGB camera \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Kim, et al. (2021)<a class="elsevierStyleCrossRef" href="#bib0445"><span class="elsevierStyleSup">33</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">5 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">The use of pulse oximetry in the assessment of acclimatization to high altitude \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Dünnwald, et al. (2021)<a class="elsevierStyleCrossRef" href="#bib0320"><span class="elsevierStyleSup">8</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Evaluación cualitativa-aceptable(Revisión bibliográfica) \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Analysis of discrepancies between pulse oximetry and arterial oxygen saturation measurements by race and ethnicity and association with organ dysfunction and mortality \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Wong, et al. (2021)<a class="elsevierStyleCrossRef" href="#bib0450"><span class="elsevierStyleSup">34</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Evaluación cualitativa-aceptable(Revisión bibliográfica) \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Accuracy of pulse oximetry in the presence of fetal hemoglobin—a systematic review \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Pritišanac, et al. (2021)<a class="elsevierStyleCrossRef" href="#bib0455"><span class="elsevierStyleSup">35</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Aceptable \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">The effects of skin pigmentation on the accuracy of pulse oximetry in measuring oxygen saturation: a systematic review and meta-analysis \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Shi, et al. (2022)<a class="elsevierStyleCrossRef" href="#bib0460"><span class="elsevierStyleSup">36</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Evaluación cualitativa-aceptable(Revisión bibliográfica) \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Springer Link \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">COVID-19: Pulse oximeters in the spotlight \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Michard, et al. (2021)<a class="elsevierStyleCrossRef" href="#bib0465"><span class="elsevierStyleSup">37</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Evaluación cualitativa-aceptable(Revisión bibliográfica) \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">National Library of Medicine \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Henna dye: a cause of erroneous pulse oximetry readings \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Bensghir, et al. (2013)<a class="elsevierStyleCrossRef" href="#bib0470"><span class="elsevierStyleSup">38</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">5 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " rowspan="4" align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">PubMed</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">A multi-wavelength opto-electronic patch sensor to effectively detect physiological changes against human skin types \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Yan, et al. (2017)<a class="elsevierStyleCrossRef" href="#bib0350"><span class="elsevierStyleSup">14</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">8 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Pulse oximetry for monitoring patients with COVID-19 at home. Potential pitfalls and practical guidance \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Luks & Swenson, (2020)<a class="elsevierStyleCrossRef" href="#bib0335"><span class="elsevierStyleSup">11</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Evaluación cualitativa-aceptable(Revisión bibliográfica) \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Racial bias in pulse oximetry measurement \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Sjoding, et al. (2020)<a class="elsevierStyleCrossRef" href="#bib0475"><span class="elsevierStyleSup">39</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">9 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Exploring the impact of pulse oximeter selection within the COVID-19 home-use pulse oximetry pathways \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Stell, et al. (2022)<a class="elsevierStyleCrossRef" href="#bib0480"><span class="elsevierStyleSup">40</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">9 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">- \t\t\t\t\t\t\n \t\t\t\t</td></tr></tbody></table> """ ] "imagenFichero" => array:1 [ 0 => "xTab3530169.png" ] ] ] ] "descripcion" => array:1 [ "es" => "<p id="spar0055" class="elsevierStyleSimplePara elsevierViewall">Selección de los artículos según base de datos</p>" ] ] 3 => array:8 [ "identificador" => "tbl0010" "etiqueta" => "Tabla 2" "tipo" => "MULTIMEDIATABLA" "mostrarFloat" => true "mostrarDisplay" => false "detalles" => array:1 [ 0 => array:3 [ "identificador" => "at2" "detalle" => "Tabla " "rol" => "short" ] ] "tabla" => array:2 [ "leyenda" => "<p id="spar0065" class="elsevierStyleSimplePara elsevierViewall">NIRS: espectroscopia en la cercanía del espectro infrarrojo.</p>" "tablatextoimagen" => array:1 [ 0 => array:2 [ "tabla" => array:1 [ 0 => """ <table border="0" frame="\n \t\t\t\t\tvoid\n \t\t\t\t" class=""><thead title="thead"><tr title="table-row"><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " colspan="4" align="center" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Resultados de la búsqueda en la literatura</th></tr><tr title="table-row"><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Autores (año), país \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Objetivo general de la investigación \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Metodología de estudio aplicada \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Principales resultados o conclusiones \t\t\t\t\t\t\n \t\t\t\t\t\t</th></tr></thead><tbody title="tbody"><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Bensghir et al.<a class="elsevierStyleCrossRef" href="#bib0470"><span class="elsevierStyleSup">38</span></a> (2013), Marruecos \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Destacar cómo la pigmentación de la piel de manos y pies del paciente influye en los resultados de la medición de la saturación de oxígeno. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Estudio de caso, consideró la presencia de henna en un paciente programado para anestesia general. Estudio centrado en uno de los factores que pueden alterar la precisión del oxímetro de pulso. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">La interferencia en la transmisión de la luz, la pigmentación oscura de la piel causada por ciertos tintes para la piel y la henna negra causan errores en las lecturas de saturación de oxígeno. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Abay y Kyriacou<a class="elsevierStyleCrossRef" href="#bib0435"><span class="elsevierStyleSup">31</span></a> (2015), Londres \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Explorar la utilización de fotopletismografía de doble longitud de onda mediante el uso de fotopletismografía pulsátil y continua para la estimación de la saturación de oxígeno arterial en postoperatorio convencional. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Aplicaron fotopletismografía de doble longitud de onda mediante el uso de fotopletismografía pulsátil y continua para la estimación de la saturación de oxígeno arterial. El sistema fue evaluado en el antebrazo de 21 voluntarios sanos durante la inducción de la oclusión venosa y la oclusión total. Se aplicaron una sonda fotopletismográfica de reflectancia y un sensor NIRS sobre el braquiorradial, se aplicaron sensores de postoperatorio en los dedos y todas las señales se adquirieron simultáneamente. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">El sistema demostró la capacidad de extraer información valiosa de las señales de fotopletismografía para una evaluación más amplia de la perfusión sanguínea de los tejidos. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Hernández-Rivera y Suaste-Gómez<a class="elsevierStyleCrossRef" href="#bib0395"><span class="elsevierStyleSup">23</span></a> (2017), México \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Obtener la señal pletismográfica teniendo en cuenta las propiedades de un polímero que cuenta con electrodos de banda necesarios para la técnica de medición de la oximetría. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Caso de estudio, se utilizó la técnica de impedanciometría en su configuración tetrapolar para obtener la señal pletismográfica del sujeto evaluado, en donde los electrodos de banda externos suministraban una corriente eléctrica y los 2 electrodos internos medían el voltaje resultante. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">La aplicación de técnica de impedanciometría en su configuración tetrapolar para obtener la señal pletismográfica del sujeto resultó más uniforme en la región censada por los electrodos internos, resultando en una medición más confiable que los convencionales. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Perillán<a class="elsevierStyleCrossRef" href="#bib0400"><span class="elsevierStyleSup">24</span></a> (2016), Chile \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Comparar los factores que afectan las mediciones de la evaluación de la saturación de oxígeno de los equipos Masimo vs. Nellcor N-200. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Se efectuó un análisis en la literatura que consideró estudios en los que aplicaron comparaciones de equipos como Masimo respecto a otros como: Nellcor N-200, Agilent Viridia M3, Masimo SET, Nellcor N-3000 Oxismart y Philips FAST. Se destacan los principales factores que influyen en la medición de la oximetría de pulso. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Los principales factores de error fueron los artefactos por movimiento, hipoperfusión, el ruido ambiental, la luz ambiental intensa, fármacos que afectan la perfusión, pigmentación de la piel, carboxihemoglobina, hemoglobina fetal, metahemoglobina, temperatura y altura. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Yan et al.<a class="elsevierStyleCrossRef" href="#bib0350"><span class="elsevierStyleSup">14</span></a> (2017), China \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Investigar la efectividad de la medición de la iluminación de múltiples longitudes de onda basada en un sensor de parche optoelectrónico contra los pigmentos de la piel utilizando una nueva arquitectura electrónica con ajuste autoadaptativo de señales. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Estudio de casos; se diseñó un protocolo para investigar el impacto de 5 tipos de piel (blanca caucásica, blanca europea, marrón claro, marrón moderado, marrón oscuro) de acuerdo con la escala cromática de Von Luschan. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Un sensor de parche optoelectrónico de múltiples longitudes de onda detecta eficazmente señales fisiológicas contra los pigmentos de la piel. Sin embargo, las diferentes ubicaciones de medición del parche optoelectrónico en la piel con vasos sanguíneos periféricos, iluminaciones de longitud de onda, artefactos de movimiento e incluso ruidos electrónicos pueden alterar los resultados de la medición de la saturación de oxígeno. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Baek et al.<a class="elsevierStyleCrossRef" href="#bib0440"><span class="elsevierStyleSup">32</span></a> (2018), Corea \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Diseñar un sensor que evite la diafonía directa en el sensor de oximetría de pulso de tipo reflectancia a través de análisis teóricos y experimentos. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Se hicieron sensores de diafonía y sensores sin diafonía y se experimentó desaturación con control de gas respiratorio en sujetos con varios colores de piel para realizar mediciones de saturación de oxígeno. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">La presencia de diafonías aumentó la probabilidad de error en la medida de la saturación de oxígeno en función del color de la piel. La mayoría de los errores en la saturación de oxígeno en los sensores convencionales se deben a diafonías ópticas que se producen en sujetos de piel oscura. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Arnaudo<a class="elsevierStyleCrossRef" href="#bib0405"><span class="elsevierStyleSup">25</span></a> (2019), Argentina \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Indagar en la literatura sobre la aplicación de nuevas tecnologías en el cuidado neonatal (NIRS) y los principios en los que se basa su funcionamiento. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Se llevó a cabo una recopilación de información relacionada con la incorporación de nuevas tecnologías al cuidado neonatal (NIRS), los principios en los que se basa su funcionamiento y los cuidados de enfermería referidos al equipamiento y al recién nacido. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">La medición del sensor varía dependiendo de las características del cráneo, las estructuras extracraneales, el flujo sanguíneo y la profundidad de la superficie cerebral. En los recién nacidos puede realizar mediciones de la oxigenación regional de órganos más profundos, como los riñones y los intestinos. También puede medir la oxigenación de los tejidos periféricos, como el antebrazo, la pantorrilla, la parte superior del brazo y el muslo. Se recomienda evitar la colocación del sensor sobre depósitos grasos, hematomas, nevos y lesiones dérmicas, ya que puede generar lecturas erráticas. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Santos et al.<a class="elsevierStyleCrossRef" href="#bib0410"><span class="elsevierStyleSup">26</span></a> (2020), México \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Conocer la variabilidad de la medición de acuerdo con el oxímetro de pulso utilizado en la práctica clínica. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Se llevó a cabo un diseño transversal. Se realizó una medición de desaturación para reducir la saturación de oxígeno al 70% mediante el control del gas respiratorio, y se evaluaron los resultados de la medición de la saturación de oxígeno; igualmente, se efectuó una comparación usando sensores con y sin prevención de diafonía. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Los 3 oxímetros presentados en este estudio (Onyx II, Hergom y SB220) exhibieron cierto grado de variabilidad. El SB220 mostró amplia variabilidad, no aconsejándose su uso clínico. El Onyx II y el Hergom mantuvieron la exactitud y la precisión. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Sjoding et al.<a class="elsevierStyleCrossRef" href="#bib0475"><span class="elsevierStyleSup">39</span></a> (2020), EE. UU. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Conocer la importancia clínica del posible sesgo racial en la medición de la oximetría de pulso. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Se analizaron medidas pareadas de oximetría de pulso de saturación de oxígeno y medidas de saturación de oxígeno arterial en gases en sangre arterial, y todas las evaluaciones se llevaron a cabo con un intervalo de 10<span class="elsevierStyleHsp" style=""></span>min entre sí. Para garantizar la medición de la saturación de oxígeno arterial se evaluaron directamente mediante cooximetría, se limitaron los análisis a medidas de gases en sangre arterial que incluían saturaciones de carboxihemoglobina y metahemoglobina. Luego se realizaron pruebas de hipoxemia oculta entre pacientes que identificaron su raza como negra o blanca. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Los pacientes de pigmentación oscura (negros) mostraron 3 veces más frecuencia de hipoxemia oculta que no fue detectada por oximetría de pulso en comparación con los pacientes de baja pigmentación (blancos). Por lo tanto, los niveles de oxígeno suplementarios pueden colocar a los pacientes con mayor pigmentación en mayor riesgo de hipoxemia. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Puentes López et al.<a class="elsevierStyleCrossRef" href="#bib0415"><span class="elsevierStyleSup">27</span></a> (2020), Colombia \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Hacer una descripción breve sobre oximetría cerebral no invasiva mediante la tecnología NIRS. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Realizaron una revisión de la literatura enfocada en la oximetría cerebral no invasiva mediante la tecnología NIRS. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">La monitorización continua cerebral mediante NIRS en enfermedades pediátricas apuesta a evitar lesiones cerebrales secundarias a hipoxia o isquemia, pero aún se necesitan para su uso rutinario estudios de gran tamaño con mejores diseños, como ensayos de control aleatorios. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Luks y Swenson<a class="elsevierStyleCrossRef" href="#bib0335"><span class="elsevierStyleSup">11</span></a> (2020), EE. UU. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Evaluar la precisión de los dispositivos de monitorización y el examen de los datos disponibles sobre el rendimiento de oxímetros de pulso económicos y sistemas basados en teléfonos inteligentes. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Revisión bibliográfica; se investigaron diferentes dispositivos disponibles y sus principios de funcionamiento, como también se analizaron las principales herramientas para evaluar la precisión de los dispositivos de monitorización. Se revisaron algunos errores potenciales con la monitorización de oximetría de pulso que podrían afectar la precisión y la implementación de un programa de monitorización. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Entre las variables que pueden afectar la exactitud de la monitorización en los resultados de la oximetría de pulso se hallaron: posición en la curva de disociación hemoglobina-oxígeno y algunas acciones que implican ventilación. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Bosio et al.<a class="elsevierStyleCrossRef" href="#bib0420"><span class="elsevierStyleSup">28</span></a> (2021), Argentina \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Comparar las mediciones de 2 saturómetros de pulso para evaluar la concordancia en pacientes con enfermedad pulmonar, utilizando la cooximetría como técnica de referencia. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Se calculó sensibilidad y especificidad de pletismografía para los pacientes que desaturan o no, según un punto de corte establecido en 93% por los gases en sangre. Se utilizaron 2 oxímetros de pulso (primera marca: saturómetro 1; y segunda marca: saturómetro 2). Se procedió a la toma de muestra de gases en sangre en la misma mano en la que se tomó la saturación. A los pacientes con requerimiento de oxígeno, este les fue retirado 20<span class="elsevierStyleHsp" style=""></span>min antes de la extracción. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">En la comparación entre los 2 oxímetros estudiados se halló que el oxímetro de pulso que utilizó solo 2 longitudes de onda no fue capaz de diferenciar la oxihemoglobina de la carboxihemoglobina, y cuando esta última se encontraba en niveles por encima de 3-4%, generaba una saturación de oxígeno que se hallaba por encima del porcentaje de oxigenoterapia hiperbárica de la sangre. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Michard et al.<a class="elsevierStyleCrossRef" href="#bib0465"><span class="elsevierStyleSup">37</span></a> (2021), Suiza \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Describir innovaciones recientes en oximetría de pulso que podrían ser útiles en pacientes con enfermedad por COVID-19. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Revisión bibliográfica. Hace referencia al uso de un sensor inalámbrico desechable para el dedo que permite la monitorización continua de saturación de oxígeno en el hogar. Un centro de comando recibe mediciones y alertas a través del teléfono inteligente del paciente. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Desde el hogar hasta la Unidad Cuidados Intensivos, las innovaciones en oximetría de pulso van más allá de la monitorización remota de saturación de oxígeno y son susceptibles de mejorar la calidad de la atención en pacientes que desarrollan insuficiencia respiratoria aguda, particularmente aquellos con COVID-19. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Kim et al.<a class="elsevierStyleCrossRef" href="#bib0445"><span class="elsevierStyleSup">33</span></a> (2021), Corea \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Medir la saturación de oxígeno a través de un método sin contacto, utilizando una cámara roja, verde, azul en un entorno de luz ambiental. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Se utilizaron las características de absorción opuestas de la hemoglobina en las longitudes de onda verde y roja. Las señales se extrajeron convirtiendo el espacio de color rojo, verde, azul de la señal extraída de imágenes faciales al espacio de color. Después de obtener cada señal extraída, se encontraron los puntos pico y valle y se tomaron los logaritmos de la relación pico/valle. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Al considerar las características de absorción opuesta de la hemoglobina en longitudes de ondas verde y roja se pudo identificar el comportamiento de la saturación de oxígeno según el entorno de luz.La señal sin procesar contuvo artefactos de movimiento y ruido que podían afectar la calidad de la señal. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Dünnwald et al.<a class="elsevierStyleCrossRef" href="#bib0320"><span class="elsevierStyleSup">8</span></a> (2021), Austria \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Proporcionar información sobre el funcionamiento de los oxímetros de pulso, los métodos de medición apropiados y los cursos de tiempo publicados de los datos de oximetría de pulso (saturación de oxígeno periférico) y la frecuencia cardíaca, registrados en reposo y ejercicio submáximo durante la exposición a diversas altitudes. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Efectuaron búsqueda bibliográfica. Se consideraron los principios básicos de funcionamiento, los obstáculos más relevantes y las posibles contramedidas para la oximetría de pulso, particularmente en personas sanas que viajan a grandes altitudes. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Existen variaciones de las medidas de oximetría de pulso y frecuencia cardíaca durante la exposición aguda y la aclimatación a grandes alturas. Los niveles de presión parcial de oxígeno disminuyen con la exposición aguda a la altitud/hipoxia y se recuperan parcialmente durante la aclimatación, con una tendencia opuesta en la frecuencia cardíaca. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Wong et al.<a class="elsevierStyleCrossRef" href="#bib0450"><span class="elsevierStyleSup">34</span></a> (2021), EE. UU. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Examinar las discrepancias raciales y étnicas entre las medidas de saturación de oxígeno y presión parcial de oxígeno y sus asociaciones con los resultados clínicos. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Se estudiaron pacientes adultos hospitalizados que recibían oxígeno suplementario y pacientes en unidades de cuidados intensivos en 178 hospitales. Se evaluaron la incidencia y los resultados clínicos de la hipoxemia oculta en todos los grupos raciales y étnicos (asiáticos, negros, hispanos o blancos). \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Se demostró que todos los grupos raciales y étnicos experimentaron discrepancias en la oximetría de pulso, lo que resultó en hipoxemia oculta; sin embargo, fue más frecuente en pacientes asiáticos, negros e hispanos que en pacientes blancos. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Pritišanac et al.<a class="elsevierStyleCrossRef" href="#bib0455"><span class="elsevierStyleSup">35</span></a> (2021), Austria \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Identificar los estudios que examinaron el efecto de la hemoglobina fetal en la monitorización de la oximetría de pulso en recién nacidos a término y prematuros, entre 24 a 42 semanas. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Revisión sistemática, incluyendo como palabras clave: recién nacido, neonato, prematuro, término, lactante, hemoglobina fetal, después del nacimiento, postnatal, oxigenación, saturación arterial de oxígeno, oximetría de pulso, saturación arterial de oxígeno y saturación de oxígeno. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Plasmaron que la presencia de la hemoglobina fetal en la sangre neonatal afecta la lectura de la saturación de oxígeno, provocando una sobreestimación de la misma. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Díaz et al.<a class="elsevierStyleCrossRef" href="#bib0345"><span class="elsevierStyleSup">13</span></a> (2022), México \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Evaluar la influencia del esmalte de uñas ecológico en la medición de la saturación periférica de oxígeno en pacientes sometidos a cirugía. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Caso clínico. Se presentaron 2 pacientes con esmalte ecológico o vegano, se les hicieron mediciones de la saturación de oxígeno con oxímetro de pulso empleando el dedo, con presencia del esmalte y con posterior retiro del mismo, previo al acto quirúrgico. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Concluyeron que la presencia de esmalte de uñas influía en los resultados de la saturación de oxígeno. En los 2 casos analizados los valores de saturación de oxígeno con esmalte partieron de un rango del 92% para la primera paciente y del 94% para la segunda, comprobándose incrementos de la saturación, luego de retirar el esmalte, a 98% para la primera paciente y 99% la segunda. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Cabanas et al.<a class="elsevierStyleCrossRef" href="#bib0425"><span class="elsevierStyleSup">29</span></a> (2022), Chile \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Resumir la literatura disponible en las principales bases de datos científicas que examinan el impacto de la pigmentación de la piel en la oximetría de pulso. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Se emplearon elementos de informes preferidos para revisiones sistemáticas y metaanálisis. Se recopilaron 41 referencias que cumplieron los criterios de inclusión establecidos. Se realizó un enfoque bibliométrico y cienciométrico para examinar los resultados de un estudio del contenido temático. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Los últimos estudios han demostrado que los oxímetros de pulso tienen algunas limitaciones con sujetos de piel oscura, especialmente en condiciones de baja saturación o hipoxemia. La melanina atenúa la longitud de onda de la luz incidente y limita la penetración al tejido subcutáneo porque absorbe mucho la luz, debido a que la piel oscura contiene más melanina, lo que da mayor incidencia en el riesgo de hipoxemia oculta. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Shi et al.<a class="elsevierStyleCrossRef" href="#bib0460"><span class="elsevierStyleSup">36</span></a> (2022), Argentina \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Evaluar los efectos de la pigmentación de la piel en la precisión de la medición de la saturación de oxígeno por oximetría de pulso en comparación con el estándar de oro, la saturación de oxígeno medida por cooximetría. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Se realizaron búsquedas en diferentes bases de datos. Se identificaron estudios que compararon los valores de saturación de oxígeno en cualquier población y ámbito de atención, utilizando cualquier tipo de oxímetro de pulso, con la saturación de oxígeno mediante cooximetría estándar, y se medió el impacto de la pigmentación de la piel o el origen étnico en la precisión de la oximetría de pulso. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">La oximetría de pulso puede sobreestimar la saturación de oxígeno en personas con piel oscura y personas cuyo origen étnico se informa como negro/afroamericano, en comparación con saturación de oxígeno, aunque la sobreestimación puede ser bastante pequeña en entornos hospitalarios. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Stell et al.<a class="elsevierStyleCrossRef" href="#bib0480"><span class="elsevierStyleSup">40</span></a> (2022), Londres \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Evaluar las características de rendimiento de 5 oxímetros de pulso portátiles y su idoneidad para su implementación en vías de oximetría de pulso de uso doméstico creadas durante la pandemia de COVID-19. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Se tomaron un total de 915 mediciones de saturación de oxígeno, junto con mediciones de un oxímetro de pulso estándar de hospital, de 50 pacientes seleccionados de las salas de respiración y la Unidad de Cuidados Intensivos. Los efectos del modelo del dispositivo y varias características de los pacientes sobre el sesgo, la probabilidad de falsos negativos y falsos positivos se evaluaron mediante análisis de regresión múltiple. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Los 5 oxímetros de pulso portátiles (Oxywatch MD300C19, Oxywatch MD300C29, PC-60F, Contec CMS50D y AM801) parecieron superar el estándar con el que fueron fabricados. El modelo del dispositivo, la saturación de oxígeno del paciente y el color de la piel del paciente fueron predictores significativos del sesgo de medición. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Barrera et al.<a class="elsevierStyleCrossRef" href="#bib0290"><span class="elsevierStyleSup">2</span></a> (2022), Argentina \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Abordar el desarrollo de un oxímetro de pulso de bajo costo, integrando la placa MAX30102, una placa Arduino UNO y un visor OLED SSD1306. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Se evaluó la concordancia entre 2 métodos cuantitativos de medición: un nuevo método de medición (oxímetro de pulso integrando con la placa MAX30102, placa Arduino UNO y un visor OLED SSD1306) con un método estándar. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Si bien el prototipo presentado realizó los cálculos correctos, no se puede garantizar que el rendimiento del sistema se mantenga a lo largo del tiempo. Para ello, se recomienda establecer una simulación para poder probar el sistema en su completitud continuamente. En cuanto a las mediciones de saturación de oxígeno, se puede garantizar que fueron confiables ya que la desviación resultó ser muy leve. \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Luna et al.<a class="elsevierStyleCrossRef" href="#bib0430"><span class="elsevierStyleSup">30</span></a> (2023), Bolivia \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Determinar las variaciones fenotípicas de hemoglobina, saturación de oxígeno, P50 y lactato en andinos bolivianos residentes entre 400, 4.000 o 5.000<span class="elsevierStyleHsp" style=""></span>m s. n. m. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Se recogió sangre venosa periférica de andinos bolivianos (n<span class="elsevierStyleHsp" style=""></span>=<span class="elsevierStyleHsp" style=""></span>124) nacidos a 4.000<span class="elsevierStyleHsp" style=""></span>m, pero radican en altitudes diferentes (400, 4.000 o 5.000<span class="elsevierStyleHsp" style=""></span>m), así como de aquellos con eritrocitosis patológicas. Adicionalmente, se recogieron muestras de europeos residentes a 4.000<span class="elsevierStyleHsp" style=""></span>m (n<span class="elsevierStyleHsp" style=""></span>=<span class="elsevierStyleHsp" style=""></span>11). Se realizaron estudios de hemograma, oximetría y gasometría. La P50 fue calculada con la fórmula de Lichtman. \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Los niveles de saturación de oxígeno reflejaron distintas variaciones a diferentes altitudes. Los habitantes andinos sanos, comparados entre ellos a distintas altitudes, reflejaron disminución en los niveles de saturación de oxígeno, al ser mayor la altitud. \t\t\t\t\t\t\n \t\t\t\t</td></tr></tbody></table> """ ] "imagenFichero" => array:1 [ 0 => "xTab3530171.png" ] ] ] ] "descripcion" => array:1 [ "es" => "<p id="spar0060" class="elsevierStyleSimplePara elsevierViewall">Resultados de la búsqueda en la literatura (factores que intervienen en la medición de la saturación de oxígeno)</p>" ] ] 4 => array:8 [ "identificador" => "tbl0015" "etiqueta" => "Tabla 3" "tipo" => "MULTIMEDIATABLA" "mostrarFloat" => true "mostrarDisplay" => false "detalles" => array:1 [ 0 => array:3 [ "identificador" => "at3" "detalle" => "Tabla " "rol" => "short" ] ] "tabla" => array:2 [ "leyenda" => "<p id="spar0075" class="elsevierStyleSimplePara elsevierViewall">PPG: fotopletismografía; SpO<span class="elsevierStyleInf">2</span>: saturación de oxígeno.</p>" "tablatextoimagen" => array:1 [ 0 => array:2 [ "tabla" => array:1 [ 0 => """ <table border="0" frame="\n \t\t\t\t\tvoid\n \t\t\t\t" class=""><thead title="thead"><tr title="table-row"><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " colspan="3" align="center" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Factores que afectan la precisión de la oximetría de pulso</th></tr><tr title="table-row"><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">SpO<span class="elsevierStyleInf">2</span>Subestimación \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">SpO<span class="elsevierStyleInf">2</span>Sobreestimación \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">PPGDesfiguración \t\t\t\t\t\t\n \t\t\t\t\t\t</th></tr></thead><tbody title="tbody"><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Perfusión periférica \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Pigmentación de la piel \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Artefactos de movimiento \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Estados de baja perfusión \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Hemoglobinopatías \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Contaminación lumínica \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Temperatura de la piel fría \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Temperatura de la piel caliente \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Frecuencia cardíaca mayor de 150<span class="elsevierStyleHsp" style=""></span>lpm \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Esmalte de uñas o tatuajes \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t"> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Interferencia electromagnética \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Hipoxemia, anemia \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t"> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Colocación del sensor \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Entrenamiento a gran altitud o hipoxia \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t"> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Tinte de henna \t\t\t\t\t\t\n \t\t\t\t</td></tr></tbody></table> """ ] "imagenFichero" => array:1 [ 0 => "xTab3530170.png" ] ] ] ] "descripcion" => array:1 [ "es" => "<p id="spar0070" class="elsevierStyleSimplePara elsevierViewall">Factores que afectan a la precisión de la pulsioximetría y a la interpretación de las lecturas</p>" ] ] ] "bibliografia" => array:2 [ "titulo" => "Bibliografía" "seccion" => array:1 [ 0 => array:2 [ "identificador" => "bibs0015" "bibliografiaReferencia" => array:56 [ 0 => array:3 [ "identificador" => "bib0285" "etiqueta" => "1" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Nivel de conocimiento del uso del oxímetro de pulso para evaluar hipoxia en presuntos casos Covid-19, en alumnos de la Universidad Norbert Wiener, Lima-2021 [Trabajo de Grado]" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:2 [ 0 => "J.G. Chochoca Vasquez" 1 => "T.C. Valle Leon" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Libro" => array:3 [ "fecha" => "2021" "editorial" => "Universidad Norbert Wiener" "editorialLocalizacion" => "Lima" ] ] ] ] ] ] 1 => array:3 [ "identificador" => "bib0290" "etiqueta" => "2" "referencia" => array:1 [ 0 => array:1 [ "referenciaCompleta" => "Barrera L, Rodríguez M, Bond R, Morales M, Encinas D. (2023). Diseño de un oxímetro de pulso. XXVIII Congreso Argentino de Ciencias de la Computación (CACIC). La Rioja, 3 al 6 de ocubre de 2022. pp. 609-617. <a target="_blank" href="http://sedici.unlp.edu.ar/handle/10915/149426">http://sedici.unlp.edu.ar/handle/10915/149426</a>" ] ] ] 2 => array:3 [ "identificador" => "bib0295" "etiqueta" => "3" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Wearable photoplethysmographic sensors—Past and present" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:4 [ 0 => "T. Tamura" 1 => "Y. Maeda" 2 => "M. Sekine" 3 => "M. Yoshida" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.3390/electronics3020282" "Revista" => array:5 [ "tituloSerie" => "Electronics." "fecha" => "2014" "volumen" => "3" "paginaInicial" => "282" "paginaFinal" => "302" ] ] ] ] ] ] 3 => array:3 [ "identificador" => "bib0300" "etiqueta" => "4" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Diseño e implementación de dispositivo para oximetría de pulso como alternativa a pulxímetros comerciales [Trabajo de Grado-Pregrado]" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:1 [ 0 => "S.E. Alfonso Guarnizo" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Libro" => array:3 [ "fecha" => "2020" "editorial" => "Universidad de los Andes" "editorialLocalizacion" => "Bogotá, Colombia" ] ] ] ] ] ] 4 => array:3 [ "identificador" => "bib0305" "etiqueta" => "5" "referencia" => array:1 [ 0 => array:1 [ "referenciaCompleta" => "Division of Industry and Consumer Education (DICE). United States Food and Drug Administration. Pulse oximeter accuracy and limitations. Silver Spring, MD: FDA Safety Communication; 2021. [consultado 12 Jul 2023]. Disponible en: <a target="_blank" href="https://www.fda.gov/news-events/fda-brief/fda-brief-fda-warns-about-limitations-and-accuracy-pulse-oximeters">https://www.fda.gov/news-events/fda-brief/fda-brief-fda-warns-about-limitations-and-accuracy-pulse-oximeters</a>" ] ] ] 5 => array:3 [ "identificador" => "bib0310" "etiqueta" => "6" "referencia" => array:1 [ 0 => array:1 [ "referenciaCompleta" => "Food and Drug Administration. Approach for improving the performance evaluation of pulse oximeter devices taking into consideration skin pigmentation, race and ethnicity discussion paper and request for feedback approach for improving the performance evaluation of pulse oximeter device. Silver Spring, MD: FDA; 2023. [consultado 24 Ago 2023]. Disponible en: <a target="_blank" href="https://www.fda.gov/medical-devices/medical-devices-news-and-events/cdrh-takes-steps-advance-further-discussions-pulse-oximeters">https://www.fda.gov/medical-devices/medical-devices-news-and-events/cdrh-takes-steps-advance-further-discussions-pulse-oximeters</a>" ] ] ] 6 => array:3 [ "identificador" => "bib0315" "etiqueta" => "7" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Factores que afectan la oximetría de pulso" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:1 [ 0 => "E.M. Rojas" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Revista" => array:6 [ "tituloSerie" => "Rev Mex Anestesiol" "fecha" => "2006" "volumen" => "29" "numero" => "S1" "paginaInicial" => "193" "paginaFinal" => "198" ] ] ] ] ] ] 7 => array:3 [ "identificador" => "bib0320" "etiqueta" => "8" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "The use of pulse oximetry in the assessment of acclimatization to high altitude" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:4 [ 0 => "T. Dünnwald" 1 => "R. Kienast" 2 => "D. Niederseer" 3 => "M. Burtscher" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.3390/s21041263" "Revista" => array:5 [ "tituloSerie" => "Sensors." "fecha" => "2021" "volumen" => "21" "paginaInicial" => "1263" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/33578839" "web" => "Medline" ] ] ] ] ] ] ] ] 8 => array:3 [ "identificador" => "bib0325" "etiqueta" => "9" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "La pulsioximetría y su aplicación en pruebas de esfuerzo máximo" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:4 [ 0 => "C. Marca Fuertes" 1 => "M. Galindo Canales" 2 => "F. Miguel-Tobal" 3 => "P. Martín Escudero" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1016/j.apunts.2010.11.002" "Revista" => array:5 [ "tituloSerie" => "Apunts Med Esport." "fecha" => "2011" "volumen" => "46" "paginaInicial" => "23" "paginaFinal" => "27" ] ] ] ] ] ] 9 => array:3 [ "identificador" => "bib0330" "etiqueta" => "10" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Métodos no invasivos para la detección de hemoglobina en sangre [Trabajo de Grado-Pregrado en Medicina]" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:1 [ 0 => "L.E. Brando de Angelis" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Libro" => array:3 [ "fecha" => "2017" "editorial" => "Universidad del Norte" "editorialLocalizacion" => "Barranquilla, Colombia" ] ] ] ] ] ] 10 => array:3 [ "identificador" => "bib0335" "etiqueta" => "11" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Pulse oximetry for monitoring patients with COVID-19 at home. Potential pitfalls and practical guidance" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:2 [ 0 => "A.M. Luks" 1 => "E.R. Swenson" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1513/AnnalsATS.202005-418FR" "Revista" => array:6 [ "tituloSerie" => "Ann Am Thorac Soc." "fecha" => "2020" "volumen" => "17" "paginaInicial" => "1040" "paginaFinal" => "1046" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/32521167" "web" => "Medline" ] ] ] ] ] ] ] ] 11 => array:3 [ "identificador" => "bib0340" "etiqueta" => "12" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Avaliação das medidas de oximetria de pulso em indivíduos sadios com esmalte de unha" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:5 [ 0 => "S. Diccini" 1 => "E.M. Pereira" 2 => "S.Y. Im" 3 => "L.Y. Shida" 4 => "A.R. Bettencourt" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1590/S0103-21002011000600009" "Revista" => array:5 [ "tituloSerie" => "Acta Paul Enferm." "fecha" => "2011" "volumen" => "24" "paginaInicial" => "784" "paginaFinal" => "788" ] ] ] ] ] ] 12 => array:3 [ "identificador" => "bib0345" "etiqueta" => "13" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Esmaltes de uñas ecológicos y medición de la saturación periférica de oxígeno en pacientes quirúrgicos. Presentación de casos" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:3 [ 0 => "J.H. Díaz" 1 => "A.M. Pérez" 2 => "Y. Díaz" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.35366/106349" "Revista" => array:5 [ "tituloSerie" => "Rev Mex Anestesiol." "fecha" => "2022" "volumen" => "45" "paginaInicial" => "285" "paginaFinal" => "288" ] ] ] ] ] ] 13 => array:3 [ "identificador" => "bib0350" "etiqueta" => "14" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "A multi-wavelength opto-electronic patch sensor to effectively detect physiological changes against human skin types" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:5 [ 0 => "L. Yan" 1 => "S. Hu" 2 => "A. Alzahrani" 3 => "S. Alharbi" 4 => "P. Blanos" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.3390/bios7020022" "Revista" => array:5 [ "tituloSerie" => "Biosensors (Basel)." "fecha" => "2017" "volumen" => "7" "paginaInicial" => "22" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/28635643" "web" => "Medline" ] ] ] ] ] ] ] ] 14 => array:3 [ "identificador" => "bib0355" "etiqueta" => "15" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Improving pulse oximetry accuracy in dark-skinned patients: Technical aspects and current regulations" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:3 [ 0 => "A.M. Cabanas" 1 => "P. Martín-Escudero" 2 => "K.H. Shelley" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1016/j.bja.2023.07.005" "Revista" => array:6 [ "tituloSerie" => "Br J Anaesth." "fecha" => "2023" "volumen" => "131" "paginaInicial" => "640" "paginaFinal" => "644" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/37544838" "web" => "Medline" ] ] ] ] ] ] ] ] 15 => array:3 [ "identificador" => "bib0360" "etiqueta" => "16" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Sistema en electrónica impreso para el monitoreo de temperatura corporal, ritmo cardíaco, frecuencia respiratoria y saturación de oxígeno en recién nacidos [Trabajo de Grado]" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:2 [ 0 => "J.C. Benavides" 1 => "J.O. González" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Libro" => array:3 [ "fecha" => "2023" "editorial" => "Universidad El Bosque" "editorialLocalizacion" => "Bogotá, Colombia" ] ] ] ] ] ] 16 => array:3 [ "identificador" => "bib0365" "etiqueta" => "17" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Oximetría de pulso" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:2 [ 0 => "H. Mejía Salas" 1 => "M. Mejía Suárez" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Revista" => array:5 [ "tituloSerie" => "RevSoc Boliv Pediatr" "fecha" => "2012" "volumen" => "51" "paginaInicial" => "149" "paginaFinal" => "155" ] ] ] ] ] ] 17 => array:3 [ "identificador" => "bib0370" "etiqueta" => "18" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Hipoxemia silenciosa y uso de oxímetro de pulso en el hogar para detección precoz de complicaciones por Covid-19: reporte de un caso y revisión de la literatura" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:2 [ 0 => "V. Berna" 1 => "L. Mendoza" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.5354/2735-7996.2021.69664" "Revista" => array:5 [ "tituloSerie" => "Rev Hosp Clin Univ Chile." "fecha" => "2021" "volumen" => "32" "paginaInicial" => "23" "paginaFinal" => "28" ] ] ] ] ] ] 18 => array:3 [ "identificador" => "bib0375" "etiqueta" => "19" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Variabilidad en la saturación de oxígeno en relación al uso de mascarilla en el personal sanitario del Hospital General Ambato-IESS" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:5 [ 0 => "P.L. Paredes Lascano" 1 => "L.A. Bravo Paredes" 2 => "D.A. Álvarez Guerrero" 3 => "K.E. Benavides Vargas" 4 => "D.S. Freire Pazmiño" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.31243/mdc.uta.v6i2.1639.2022" "Revista" => array:5 [ "tituloSerie" => "Mediciencias UTA." "fecha" => "2022" "volumen" => "6" "paginaInicial" => "84" "paginaFinal" => "94" ] ] ] ] ] ] 19 => array:3 [ "identificador" => "bib0380" "etiqueta" => "20" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Diseño y construcción de un oxímetro de pulso portátil [Trabajo de Grado]" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:3 [ 0 => "P. Salgado" 1 => "J. Ronquillo" 2 => "O. Alvarado" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Libro" => array:3 [ "fecha" => "2013" "editorial" => "Universidad del Azuay" "editorialLocalizacion" => "Cuenca, Ecuador" ] ] ] ] ] ] 20 => array:3 [ "identificador" => "bib0385" "etiqueta" => "21" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Pulse oximetry: Fundamentals and technology update" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:3 [ 0 => "M. Nitzan" 1 => "A. Romem" 2 => "R. Koppel" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.2147/MDER.S47319" "Revista" => array:4 [ "tituloSerie" => "Med Devices (Auckl)" "fecha" => "2014" "paginaInicial" => "231" "paginaFinal" => "239" ] ] ] ] ] ] 21 => array:3 [ "identificador" => "bib0390" "etiqueta" => "22" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "La declaración PRISMA 2020: una guía actualizada para reportar revisiones sistemáticas" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:1 [ 0 => "A. Ciapponi" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.51987/evidencia.v24i4.6960" "Revista" => array:3 [ "tituloSerie" => "Evidencia." "fecha" => "2021" "volumen" => "24" ] ] ] ] ] ] 22 => array:3 [ "identificador" => "bib0395" "etiqueta" => "23" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Pletismografía de impedancia basada en electrodos de Polipirrol para el sensado de la frecuencia cardiaca" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:2 [ 0 => "D. Hernández-Rivera" 1 => "E. Suaste-Gómez" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Revista" => array:5 [ "tituloSerie" => "Memorias XXXIX Congreso Nacional de Ingeniería Biomédica" "fecha" => "2017" "volumen" => "3" "paginaInicial" => "100" "paginaFinal" => "103" ] ] ] ] ] ] 23 => array:3 [ "identificador" => "bib0400" "etiqueta" => "24" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Oximetría de pulso: ¿cuál es el equipo ideal?" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:1 [ 0 => "J.A. Perillán" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.51451/np.v11i2.306" "Revista" => array:5 [ "tituloSerie" => "Neumol Pediatr." "fecha" => "2016" "volumen" => "11" "paginaInicial" => "81" "paginaFinal" => "84" ] ] ] ] ] ] 24 => array:3 [ "identificador" => "bib0405" "etiqueta" => "25" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Aplicación de nuevas tecnologías en el cuidado neonatal: monitorización de la saturación regional de oxígeno" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:1 [ 0 => "P.D. Arnaudo" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Revista" => array:5 [ "tituloSerie" => "Enfermería Neonatal." "fecha" => "2019" "volumen" => "31" "paginaInicial" => "14" "paginaFinal" => "21" ] ] ] ] ] ] 25 => array:3 [ "identificador" => "bib0410" "etiqueta" => "26" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Diferencias en la saturación parcial de oxígeno obtenida de tres oxímetros de pulso" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:6 [ 0 => "L.E. Santos" 1 => "M.V. Ureña" 2 => "J. Quevedo" 3 => "N.I. Colín" 4 => "J.P. Durán" 5 => "P. Soto" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.24875/RMIMSS.M19000062" "Revista" => array:6 [ "tituloSerie" => "Rev Med Inst Mex Seguro Soc" "fecha" => "2020" "volumen" => "58" "paginaInicial" => "385" "paginaFinal" => "393" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/34543543" "web" => "Medline" ] ] ] ] ] ] ] ] 26 => array:3 [ "identificador" => "bib0415" "etiqueta" => "27" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "¿La espectroscopia de infrarrojo cercano (NIRS) podría ser útil en el monitoreo hemodinámico del paciente critico pediátrico?" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:5 [ 0 => "S.P. Puentes López" 1 => "A.C. Zárate Vergara" 2 => "I.S. Tirado Pérez" 3 => "Y. García Orozco" 4 => "Y. Ariza Álvarez" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.18041/2390-0512/biociencias.1.6363" "Revista" => array:5 [ "tituloSerie" => "Biociencias." "fecha" => "2020" "volumen" => "15" "paginaInicial" => "73" "paginaFinal" => "78" ] ] ] ] ] ] 27 => array:3 [ "identificador" => "bib0420" "etiqueta" => "28" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Estudio retrospectivo para describir la concordancia entre dos oxímetros de pulso" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:6 [ 0 => "M. Bosio" 1 => "G. Ernst" 2 => "J. Pascua" 3 => "P. Oyamburu" 4 => "F. di Tullio" 5 => "A. Salvado" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.31954/RFEM/202101/0027-0030" "Revista" => array:5 [ "tituloSerie" => "Front Med." "fecha" => "2021" "volumen" => "16" "paginaInicial" => "27" "paginaFinal" => "74" ] ] ] ] ] ] 28 => array:3 [ "identificador" => "bib0425" "etiqueta" => "29" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Skin pigmentation influence on pulse oximetry accuracy: A systematic review and bibliometric analysis" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:5 [ 0 => "A.M. Cabanas" 1 => "M. Fuentes-Guajardo" 2 => "K. Latorre" 3 => "D. León" 4 => "P. Martín-Escudero" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.3390/s22093402" "Revista" => array:5 [ "tituloSerie" => "Sensors (Basel)." "fecha" => "2022" "volumen" => "22" "paginaInicial" => "3402" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/35591092" "web" => "Medline" ] ] ] ] ] ] ] ] 29 => array:3 [ "identificador" => "bib0430" "etiqueta" => "30" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Variaciones fenotípicas relacionadas a respuestas hipóxicas entre andinos a distintas altitudes" "autores" => array:1 [ 0 => array:2 [ "etal" => true "autores" => array:6 [ 0 => "J. Luna" 1 => "E. Cayo" 2 => "T. Quispe" 3 => "J.C. Valencia" 4 => "D. Patón" 5 => "L.F. Mamani" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Revista" => array:5 [ "tituloSerie" => "Rev Med La Paz." "fecha" => "2023" "volumen" => "29" "paginaInicial" => "20" "paginaFinal" => "26" ] ] ] ] ] ] 30 => array:3 [ "identificador" => "bib0435" "etiqueta" => "31" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Reflectance photoplethysmography as noninvasive monitoring of tissue blood perfusion" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:2 [ 0 => "T.Y. Abay" 1 => "P.A. Kyriacou" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1109/TBME.2015.2417863" "Revista" => array:6 [ "tituloSerie" => "IEEE Trans Biomed Eng." "fecha" => "2015" "volumen" => "62" "paginaInicial" => "2187" "paginaFinal" => "2195" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/25838515" "web" => "Medline" ] ] ] ] ] ] ] ] 31 => array:3 [ "identificador" => "bib0440" "etiqueta" => "32" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "The effect of optical crosstalk on accuracy of reflectance-type pulse oximeter for mobile healthcare" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:3 [ 0 => "H.J. Baek" 1 => "J. Shin" 2 => "J. Cho" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1155/2018/3521738" "Revista" => array:5 [ "tituloSerie" => "J Healthc Eng." "fecha" => "2018" "volumen" => "2018" "paginaInicial" => "3521738" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/30420912" "web" => "Medline" ] ] ] ] ] ] ] ] 32 => array:3 [ "identificador" => "bib0445" "etiqueta" => "33" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Non-contact oxygen saturation measurement using YCgCr color space with an RGB camera" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:4 [ 0 => "N.H. Kim" 1 => "S.G. Yu" 2 => "S.E. Kim" 3 => "E.C. Lee" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.3390/s21186120" "Revista" => array:5 [ "tituloSerie" => "Sensors." "fecha" => "2021" "volumen" => "21" "paginaInicial" => "6120" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/34577326" "web" => "Medline" ] ] ] ] ] ] ] ] 33 => array:3 [ "identificador" => "bib0450" "etiqueta" => "34" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Analysis of discrepancies between pulse oximetry and arterial oxygen saturation measurements by race and ethnicity and association with organ dysfunction and mortality" "autores" => array:1 [ 0 => array:2 [ "etal" => true "autores" => array:6 [ 0 => "A.K. Wong" 1 => "M. Charpignon" 2 => "H. Kim" 3 => "C. Josef" 4 => "A.A. de Hond" 5 => "J.J. Fojas" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1001/jamanetworkopen.2021.31674" "Revista" => array:5 [ "tituloSerie" => "JAMA Netw Open." "fecha" => "2021" "volumen" => "4" "paginaInicial" => "e2131674" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/34730820" "web" => "Medline" ] ] ] ] ] ] ] ] 34 => array:3 [ "identificador" => "bib0455" "etiqueta" => "35" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Accuracy of pulse oximetry in the presence of fetal hemoglobin—A systematic review" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:4 [ 0 => "E. Pritišanac" 1 => "B. Urlesberger" 2 => "B. Schwaberger" 3 => "G. Pichler" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.3390/children8050361" "Revista" => array:5 [ "tituloSerie" => "Children." "fecha" => "2021" "volumen" => "8" "paginaInicial" => "361" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/33946236" "web" => "Medline" ] ] ] ] ] ] ] ] 35 => array:3 [ "identificador" => "bib0460" "etiqueta" => "36" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "The accuracy of pulse oximetry in measuring oxygen saturation by levels of skin pigmentation: A systematic review and meta-analysis" "autores" => array:1 [ 0 => array:2 [ "etal" => true "autores" => array:6 [ 0 => "C. Shi" 1 => "M. Goodall" 2 => "J. Dumville" 3 => "J. Hill" 4 => "G. Norman" 5 => "O. Hamer" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1101/2022.02.16.22271062" "Revista" => array:5 [ "tituloSerie" => "BMC Med." "fecha" => "2022" "volumen" => "20" "paginaInicial" => "267" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/35971142" "web" => "Medline" ] ] ] ] ] ] ] ] 36 => array:3 [ "identificador" => "bib0465" "etiqueta" => "37" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "COVID-19: Pulse oximeters in the spotlight" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:3 [ 0 => "F. Michard" 1 => "K. Shelley" 2 => "E. L’Her" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1007/s10877-020-00550-7" "Revista" => array:6 [ "tituloSerie" => "J Clin Monit Comput." "fecha" => "2021" "volumen" => "35" "paginaInicial" => "11" "paginaFinal" => "14" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/32578070" "web" => "Medline" ] ] ] ] ] ] ] ] 37 => array:3 [ "identificador" => "bib0470" "etiqueta" => "38" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Henna dye: A cause of erroneous pulse oximetry readings" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:6 [ 0 => "M. Bensghir" 1 => "A. Houba" 2 => "J. El Hila" 3 => "R. Ahtil" 4 => "H. Azendour" 5 => "N.D. Kamili" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.4103/1658-354X. 121052" "Revista" => array:5 [ "tituloSerie" => "Saudi J Anaesth." "fecha" => "2013" "volumen" => "7" "paginaInicial" => "474" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/24348305" "web" => "Medline" ] ] ] ] ] ] ] ] 38 => array:3 [ "identificador" => "bib0475" "etiqueta" => "39" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Racial bias in pulse oximetry measurement" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:5 [ 0 => "M.W. Sjoding" 1 => "R.P. Dickson" 2 => "T.J. Iwashyna" 3 => "S.E. Gay" 4 => "T.S. Valley" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1056/NEJMc2029240" "Revista" => array:6 [ "tituloSerie" => "N Engl J Med." "fecha" => "2020" "volumen" => "383" "paginaInicial" => "2477" "paginaFinal" => "2478" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/33326721" "web" => "Medline" ] ] ] ] ] ] ] ] 39 => array:3 [ "identificador" => "bib0480" "etiqueta" => "40" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Exploring the impact of pulse oximeter selection within the COVID-19 home-use pulse oximetry pathways" "autores" => array:1 [ 0 => array:2 [ "etal" => true "autores" => array:6 [ 0 => "D. Stell" 1 => "J.J. Noble" 2 => "R.H. Kay" 3 => "M.T. Kwong" 4 => "M.J. Jeffryes" 5 => "L. Johnston" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1136/bmjresp-2021-001159" "Revista" => array:5 [ "tituloSerie" => "BMJ Open Respir Res." "fecha" => "2022" "volumen" => "9" "paginaInicial" => "e001159" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/35140169" "web" => "Medline" ] ] ] ] ] ] ] ] 40 => array:3 [ "identificador" => "bib0485" "etiqueta" => "41" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "The Newcastle-Ottawa Scale (NOS) for assessing the quality of nonrandomised studies in meta-analyses" "autores" => array:1 [ 0 => array:2 [ "etal" => true "autores" => array:6 [ 0 => "G.A. Wells" 1 => "B. Shea" 2 => "D. O’Connell" 3 => "J. Peterson" 4 => "V. Welch" 5 => "M. Losos" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Libro" => array:3 [ "fecha" => "2021" "editorial" => "The Ottawa Hospital Research Institute" "editorialLocalizacion" => "Ottawa, Ontario" ] ] ] ] ] ] 41 => array:3 [ "identificador" => "bib0490" "etiqueta" => "42" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Appendix A: Guide to the contents of a Cochrane Methodology protocol and review" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:5 [ 0 => "M. Clarke" 1 => "A.D. Oxman" 2 => "E. Paulsen" 3 => "J.P. Higgins" 4 => "S. Green" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Libro" => array:5 [ "fecha" => "2011" "paginaInicial" => "126" "paginaFinal" => "130" "editorial" => "The Cochrane Collaboration" "editorialLocalizacion" => "London" ] ] ] ] ] ] 42 => array:3 [ "identificador" => "bib0495" "etiqueta" => "43" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Man's respiratory response during and after acclimatization to high altitude" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:2 [ 0 => "H. Rahn" 1 => "A.B. Otis" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1152/ajplegacy.1949.157.3.445" "Revista" => array:5 [ "tituloSerie" => "Am J Physiol." "fecha" => "1949" "volumen" => "157" "paginaInicial" => "445" "paginaFinal" => "462" ] ] ] ] ] ] 43 => array:3 [ "identificador" => "bib0500" "etiqueta" => "44" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Adaptation to high altitude" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:2 [ 0 => "C. Lenfant" 1 => "K. Sullivan" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1056/NEJM197106102842305" "Revista" => array:6 [ "tituloSerie" => "N Engl J Med" "fecha" => "1971" "volumen" => "284" "paginaInicial" => "1298" "paginaFinal" => "1309" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/4930601" "web" => "Medline" ] ] ] ] ] ] ] ] 44 => array:3 [ "identificador" => "bib0505" "etiqueta" => "45" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Time course of augmentation and depression of hypoxic ventilatory responses at altitude" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:3 [ 0 => "M. Sato" 1 => "J.W. Severinghaus" 2 => "P. Bickler" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1152/jappl.1994.77.1.313" "Revista" => array:6 [ "tituloSerie" => "J Appl Physiol (1985)" "fecha" => "1994" "volumen" => "77" "paginaInicial" => "313" "paginaFinal" => "316" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/7961252" "web" => "Medline" ] ] ] ] ] ] ] ] 45 => array:3 [ "identificador" => "bib0510" "etiqueta" => "46" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Increased exercise SaO2 independent of ventilatory acclimatization at 4,300<span class="elsevierStyleHsp" style=""></span>m" "autores" => array:1 [ 0 => array:2 [ "etal" => true "autores" => array:6 [ 0 => "P.R. Bender" 1 => "R.E. McCullough" 2 => "R.G. McCullough" 3 => "S.Y. Huang" 4 => "P.D. Wagner" 5 => "A. Cymerman" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1152/jappl.1989.66.6.2733" "Revista" => array:6 [ "tituloSerie" => "J Appl Physiol (1985)" "fecha" => "1989" "volumen" => "66" "paginaInicial" => "2733" "paginaFinal" => "2738" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/2745337" "web" => "Medline" ] ] ] ] ] ] ] ] 46 => array:3 [ "identificador" => "bib0515" "etiqueta" => "47" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Registrador portátil de sPO<span class="elsevierStyleInf">2</span> [Trabajo de fin de Grado]" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:1 [ 0 => "L. Barea Cañas" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Libro" => array:3 [ "fecha" => "2023" "editorial" => "Universidad de Medicina" "editorialLocalizacion" => "Valladolid" ] ] ] ] ] ] 47 => array:3 [ "identificador" => "bib0520" "etiqueta" => "48" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Monitoreo de señales biológicas basado en materiales poliméricos y piezoeléctricos [Trabajo de Grado]" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:1 [ 0 => "D. Hernández Rivera" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Libro" => array:3 [ "fecha" => "2014" "editorial" => "Centro de Investigación y de Estudios Avanzados del Instintuto Politécnico Nacional" "editorialLocalizacion" => "Ciudad de México" ] ] ] ] ] ] 48 => array:3 [ "identificador" => "bib0525" "etiqueta" => "49" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Pulse oximetry" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:1 [ 0 => "A. Jubran" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1186/cc341" "Revista" => array:6 [ "tituloSerie" => "Crit Care" "fecha" => "1999" "volumen" => "3" "paginaInicial" => "R11" "paginaFinal" => "R17" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/11094477" "web" => "Medline" ] ] ] ] ] ] ] ] 49 => array:3 [ "identificador" => "bib0530" "etiqueta" => "50" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Pulse oximetry: Beyond SpO2" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:1 [ 0 => "D.R. Hess" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.4187/respcare.05208" "Revista" => array:6 [ "tituloSerie" => "Respir Care" "fecha" => "2016" "volumen" => "61" "paginaInicial" => "1671" "paginaFinal" => "1680" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/27899542" "web" => "Medline" ] ] ] ] ] ] ] ] 50 => array:3 [ "identificador" => "bib0535" "etiqueta" => "51" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Saturación de oxígeno en niños y niñas escolares sanos de 5 a 12 años en escuelas de educación básica ubicadas a la altura de 2880 a 3000 metros en la ciudad de Quito en el período de marzo-mayo 2015 [Trabajo de Grado]" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:2 [ 0 => "J.A. Novillo" 1 => "A.G. Mata" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Libro" => array:3 [ "fecha" => "2015" "editorial" => "Pontificia Universidad Católica del Ecuador" "editorialLocalizacion" => "Quito" ] ] ] ] ] ] 51 => array:3 [ "identificador" => "bib0540" "etiqueta" => "52" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Effects of skin pigmentation on pulse oxiemter accuracy at low saturation" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:3 [ 0 => "P.E. Bickler" 1 => "J. Feiner" 2 => "R.J.W. Severinghaus" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1097/00000542-200504000-00004" "Revista" => array:6 [ "tituloSerie" => "Anesthesiology." "fecha" => "2005" "volumen" => "102" "paginaInicial" => "715" "paginaFinal" => "719" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/15791098" "web" => "Medline" ] ] ] ] ] ] ] ] 52 => array:3 [ "identificador" => "bib0545" "etiqueta" => "53" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Effect of nail polish and henna on oxygen saturation determined by pulse oximetry in healthy young adult females" "autores" => array:1 [ 0 => array:2 [ "etal" => true "autores" => array:6 [ 0 => "H.S. Çiçek" 1 => "S. Gümüş" 2 => "Ö. Deniz" 3 => "Ş. Yildiz" 4 => "C.H. Açikel" 5 => "E. Çakir" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1136/emj.2010.096073" "Revista" => array:6 [ "tituloSerie" => "Emerg Med J" "fecha" => "2011" "volumen" => "28" "paginaInicial" => "783" "paginaFinal" => "785" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/20923819" "web" => "Medline" ] ] ] ] ] ] ] ] 53 => array:3 [ "identificador" => "bib0550" "etiqueta" => "54" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Cuantificación de saturación periférica de oxígeno sin contacto mediante espectroscopia de reflectancia difusa en el rango VIS-NIR [Trabajo de Grado]" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:1 [ 0 => "F. Miranda Casasola" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Libro" => array:3 [ "fecha" => "2023" "editorial" => "Universidad Nacional Autónoma de México" "editorialLocalizacion" => "Ciudad de México" ] ] ] ] ] ] 54 => array:3 [ "identificador" => "bib0555" "etiqueta" => "55" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Monitoreo respiratorio. Oximetría del pulso" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:1 [ 0 => "J.P.A. Ortega-García" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Revista" => array:6 [ "tituloSerie" => "Rev Mex Anestesiol" "fecha" => "2013" "volumen" => "36" "numero" => "Supl 1" "paginaInicial" => "S104" "paginaFinal" => "S105" ] ] ] ] ] ] 55 => array:3 [ "identificador" => "bib0560" "etiqueta" => "56" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "La oximetría en registro continuo en el esfuerzo máximo en distintas especialidades deportivas [Tesis doctoral]" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:1 [ 0 => "M.P. Martín-Escudero" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Libro" => array:3 [ "fecha" => "2003" "editorial" => "Universidad Complutense de Madrid" "editorialLocalizacion" => "Madrid" ] ] ] ] ] ] ] ] ] ] "agradecimientos" => array:1 [ 0 => array:4 [ "identificador" => "xack743267" "titulo" => "Agradecimientos" "texto" => "<p id="par0270" class="elsevierStylePara elsevierViewall">El artículo forma parte del Proyecto de Investigación FONIS SA2210178, año 2022, Evaluación de la precisión de la oximetría de pulso en función de la concentración de melanina en la piel.</p>" "vista" => "all" ] ] ] "idiomaDefecto" => "es" "url" => "/00142565/0000022400000005/v2_202405080641/S0014256524000638/v2_202405080641/es/main.assets" "Apartado" => array:4 [ "identificador" => "1030" "tipo" => "SECCION" "es" => array:2 [ "titulo" => "Revisión" "idiomaDefecto" => true ] "idiomaDefecto" => "es" ] "PDF" => "https://static.elsevier.es/multimedia/00142565/0000022400000005/v2_202405080641/S0014256524000638/v2_202405080641/es/main.pdf?idApp=WRCEE&text.app=https://revclinesp.es/" "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S0014256524000638?idApp=WRCEE" ]
año/Mes | Html | Total | |
---|---|---|---|
2024 Octubre | 2 | 0 | 2 |
2024 Septiembre | 1 | 0 | 1 |
2024 Agosto | 1 | 2 | 3 |
2024 Julio | 2 | 0 | 2 |
2024 Junio | 3 | 2 | 5 |
2024 Mayo | 21 | 12 | 33 |